Service Configuration Settings
The scale is configured from the factory to match the specified settings for each unit, as defined by the product
specifications and sales brochure. Modification of the settings can be accomplished by altering user configuration
settings P11-P19. Caution: Configuration of the settings of your scale should be carried out by a trained service
technician to ensure proper operation and accuracy. Configuration is not covered under warranty.
Setup Controls
Hold
Print
kg
Moves flashing digit
Config
Available
Menu
Settings
P11
0,1,2
P12
0-7
P13
00 - 15
P14
0, 1, 2
P15
0, 1, 2
P16
0-8
P17
0-3
P18
0 - 9
P19
0-9
Units
Change flashing digits
Default
Definition
Units key
2
configuration
7
Power-on zero
range
03
Zero button
range
0
Signal within
power on zero
point range
1
Signal not
within power on
zero point
6
Zero tracking
3
Data Filter
9
Weight stability
1
Overload range
Full scale
Tare
kg
Saves data and move down
to next Parameter setting
.
Detailed Setting
0 = KG only ; 1 = LB only ; 2 = units key active KG
and LB
0 = +/- 1% ; 1 = +/- 2% ; 2 = +/- 5% ; 3 = +/- 10% ; 4 =
+/- 20% ; 5 = +/- 50% ; 6 = +/- 100% ; 7 = no limitation
(00) = +/- 1%
(06) = +/- 20%
(01 )= +/- 2%
(07) = +/-no limit
(02 )= +/- 3%
(08 )= + 1%
(03 )= +/- 4%
(09) = + 2%
(04) =+/- 5%
(10 )= + 3%
(05 )= +/- 10%
(11) = + 4%
0 = current weight ; 1 = calibration zero ; 2 = power off
zero point
0 = current weight ; 1 = calibration zero ; 2 = power off
zero point ; 3 = continuously display "0
0 = 0d AZT off ; 1 = +/- 0.25d, 2 = +/- 0.5d ; 3 = +/- 1d;
4= +/- 1.5d ;5= +/- 2d ;6 = +/- 3d; 7= +/- 4d; 8 = +/- 5d
0 = very weak; 1 = weak; 2 = standard; 3 = strong
0 = +/- 0.5d ; 1= +/- 1d; 2= +/- 1.5d ; 3= +/- 2d; 4= +/-
3d ;5= +/- 4d ;6 = +/- 5d; 7= +/- 6d; 8 = +/- 7d; 9 = +/-
8d
0 = 0% ; 1 = +9d ; 2 = 101% ; 3 = 102% ; 4 = 405% ;
5 = 110% ; 6 = 120% ; 7 = 150%; 8 = 200%; 9 = no
limitation
On/Off
Zero
Exits setup mode
(12) = + 5%
(13) = + 10%
(14) = + 20%
(25) = + no limit
- - - - -
"