Quick Setup Guide TDL GPro™ 500 English GPro™ 500 Quick Setup Guide Content 1 Preparation Packing content Site Requirements Ambient operating conditions What you also need Optional accesssories 2 Before the installation Flange placement Flange requirements Flow conditions at measuring point Purging Purging with ...
Quick Setup Guide TDL GPro™ 500 2 Before the installation Flange placement 60 cm (23.6") The TDL head should be easily accessible. A person should be able to stand in front of it and adjust the M16 fixing bolts using two standard spanners. There should be at least 60 cm free space measured from the flange fixed to the stack and outwards as shown below.
Quick Setup Guide TDL GPro™ 500 DN100 DN50/PN25 DN65/PN25 100 mm 100 mm ANSI 4" ANSI 2"/300 lbs ANSI 2.5"/300 lbs 4" 4" Flow conditions at When deciding the placement of the GPro™ 500 TDL in the process, we recom- measuring point mend a minimum of 5 stack diameters of straight duct before and 3 stack di- ameters of straight duct after the point of measure.
Quick Setup Guide TDL GPro™ 500 Purging 0…1 L /min 0…10 L /min Pressure regulator Rotameter 1 Purge gas inlet for instrument side (6 mm or ¼" tube fitting). 2 Purge gas inlet for process side (must have a check valve). 3 Purge gas outlet for instrument side (6 mm or ¼"...
Quick Setup Guide TDL GPro™ 500 Purging with t hermal barrier Purging configuration 1 Purge gas inlet for thermal barrier (6 mm or ¼" tube fitting) 2 Purge gas outlet for thermal barrier (6 mm or ¼" tube fitting) 3 Purge gas inlet for process side (Must have a check valve) 4 Mandatory check valve (to be supplied by the user) WARNING The purge gas for the thermal barrier must always be turned on...
Page 11
Quick Setup Guide TDL GPro™ 500 Grounding and w iring ( ATEX) External earth point for > 4 mm cables (M6 × 12) Exernal earth point.
Quick Setup Guide TDL GPro™ 500 Activeanaloginputs(ATEXVersion) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Looppoweredanaloginputs(ATEX) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 Activeanaloginputs(FMVersion) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Looppoweredanaloginputs(FMVersion) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 GPro™ 500 cables Signal Description Cable no. Color Junction Box Pin no Pin no Power + 24 V Power 24 V, 5 W GND (Power) Blue RS 485 A Green Interface M400 RS 485 B Yellow (RS 485) RS 485 GND...
Page 18
Quick Setup Guide TDL GPro™ 500 Pin 1 Pin 4 Connections on motherboard in the sensor head. Note: T hesensorheadcoverofATEXversionshouldneverbeen opened,asthiswillinvalidatetheATEXcertification. Pin 1 Pin 12 Connections on IO board in the sensor head. WARNING All openings have to be closed with certified cable glands or blocking plugs of the same degree of certification as the GPro™...
If a special parameter is chosen, only this parameter will be recognized and accepted on the transmitter. Calibration The GPro 500 is calibrated at the factory and does not require calibration at installation & Startup. TDLInstallation (path: Quick Setup / TDL / Installation) While in measurement mode press the key [MENU] .
Page 21
Pressure=External – hPa – mmHg – mbar – psi – kPa In general, METTLER TOLEDO recommends the use of absolute pressure trans- 20.9 %V O ducers for more accurate signal compensation over a broad pressure range. 20.9 %Trm If, however, small pressure variations around atmospheric pressure are to be 4 mA = 940.0 mbar...
Page 22
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C If Fixed compensation is selected, the fixed temperature value with which the 20.9 %V O measurement signal will be calculated has to be keyed in manually. For the fixed 20.9 %Trm temperature, only °C can be used.
Quick Setup Guide TDL GPro™ 500 Setting the correct The flow rate of the purging will affect the effective path length and conse- process side purging quently the measurement value. Therefore the following procedure should be used. Start with a very high flow rate and gradually decrease it.
Quick Setup Guide TDL GPro™ 500 4 VerificationandMaintenance One-pointcalibrationforTDLgassensors Enter calibration mode as described in section 7.1 “Enter Calibration Mode”. 20.9 %V O 25.0 A one-point calibration of gas sensors is always a slope (i.e. with air) calibra- °C tion. A one point slope calibration is done in air or any other calibration gas with Calibrate Sensor Channel B TDL defined gas concentration.
Quick Setup Guide TDL GPro™ 500 Enter the concentration value of the sample then press the [ENTER] key to start 12.1 %V O the calculation of the calibration results. 25.0 °C B Point1 = 56.90 %sat = 57.1 %air After the calibration the slope ”S” is displayed. 12.1 %V O 25.0...
Quick Setup Guide TDL GPro™ 500 5 Error messages Message Comment No sensor on channel 3 The M400 is unable to detect any of the ISM sensor(s) it can identify. If no sensor is found it will disply the message NO SENSOR DETECTED Signal Processing Failed Fitting of the line profiles failed.
Page 27
– Check with the MT-TDL software and the Ethernet port if the system is running correctly. – If timeout still occurs after 60 s, send unit back to METTLER TOLEDO. Send unit back to METTLER TOLEDO Fault Software error Send unit back to METTLER TOLEDO Fault System error Clean corner cube and process window.
Page 29
Quick Setup Guide TDL GPro™ 500 German GPro™ 500 Quick Setup-Leitfaden Inhaltsverzeichnis 1 Vorbereitung Lieferumfang Anforderungen vor Ort Umgebungs-bedingungen Außerdem erforderlich Optionales Zubehör 2 Vor der Installation Montageort für Flansch Anforderungen an den Flansch Strömungs-verhältnisse an der Messstelle Spülen Spülen mit Wärmeschutz Erdung und Verdrahtung (ATEX) Aktive Analogeingänge (ATEX-Ausführung) Schleifengespeiste Analogeingänge (ATEX-Ausführung)
Quick Setup Guide TDL GPro™ 500 2 Vor der Installation Montageort für Flansch 60 cm (23.6") Der TDL-Kopf muss leicht zugänglich sein. Eine Person muss vor dem Sensor stehen können und mit zwei normalen Gabelschlüsseln die M16 Befestigungs- schrauben erreichen. Um den quer in die Leitung eingebauten Flansch herum müssen mindestens 60 cm Platz sein, wie unten dargestellt.
Quick Setup Guide TDL GPro™ 500 DN100 DN50/PN25 DN65/PN25 100 mm 100 mm ANSI 4" ANSI 2"/300 lbs ANSI 2.5"/300 lbs 4" 4" Strömungs- Bei der Wahl des Einbauorts des GPro™ 500 TDL in den Prozess ist darauf zu Bei der Wahl des Einbauorts des GPro™ 500 TDL in den Prozess ist darauf zu verhältnisse achten, dass vor der Messstelle eine gerade Einlaufstrecke von mindestens dem an der Messstelle...
Quick Setup Guide TDL GPro™ 500 Spülen 0 −1 l/min 0 −10 l/min Druckregler Rotameter 1 Spülgaseinlass Geräteseite (6 mm- oder ¼"-Anschluss). 2 Spülgaseinlass Prozessseite (Absperrventil erforderlich). 3 Spülgasauslass Geräteseite (6 mm- oder ¼"-Anschluss). 4 Prozessgasfluss. 5 Bereich der Grenzen der effektiven optischen Weglänge. WARNUNG Bevor der Prozess gestartet wird, ist der Spülgasfluss immer auf maximalen Durchfluss einzustellen.
Quick Setup Guide TDL GPro™ 500 Spülen mit Wärmeschutz Konfiguration beim Spülen 1 Spülgaseinlass in die thermische Barriere (6 mm- oder ¼"-Anschluss) 2 Spülgasauslass aus der thermischen Barriere (6 mm- oder ¼"-Anschluss) 3 Spülgaseinlass Prozessseite (erfordert ein Absperrventil) 4 Absperrventil vorgeschrieben (muss vom Benutzer bereitgestellt werden) WARNUNG Das Spülgas für die thermische Barriere muss immer aktiviert sein, wenn der Prozessgasstrom aktiviert ist, um den Sensorkopf vor...
Quick Setup Guide TDL GPro™ 500 AktiveAnalogeingänge(ATEX-Ausführung) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 SchleifengespeisteAnalogeingänge(ATEX-Ausführung) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND M400 TB4 RS-485 B RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 AktiveAnalogeingänge(FM-Ausführung) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 SchleifengespeisteAnalogeingänge(FM-Ausführung) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 GPro™ 500 Kabel Signal Beschreibung Kabel-Nr. Farbe Anschlussbox Pin-Nr. Pin-Nr. Stromversorgung + 24 V Stromversorgung 24 V, GND (Stromver- Blau sorgung) RS485 A Grün Schnittstelle M400 RS485 B Gelb (RS485) RS485 GND Braun 4 ... 20 mA pos Violett Stromeingang Temperatur...
Page 44
Quick Setup Guide TDL GPro™ 500 Pin 1 Pin 4 Anschlüsse an der Hauptplatine im Sensorkopf Hinweis: DieAbdeckungdesSensorkopfsdarfbeiderATEX- Ausführung keinesfalls geöffnet werden, da sonst dieATEX-ZertifizierungihreGültigkeitverliert. Pin 1 Pin 12 Anschlüsse auf E / A-Platine im Sensorkopf WARNUNG Alle Öffnungen sind mit zugelassenen Kabelverschraubungen oder Verschlussstopfen zu verschließen, die über die gleichen Zertifizierungen verfügen wie der GPro™...
Sensoren erkannt und akzeptiert werden. Wird ein bestimmter Parameter aus- gewählt, wird nur dieser Parameter vom Transmitter erkannt und akzeptiert. Kalibrierung Der GPro 500 ist werksseitig kalibriert und erfordert bei Installation und Inbe- triebnahme keine Kalibrierung. TDLInstallation (Pfad: Quick Setup / TDL / Installation) Drücken Sie im Messmodus die Taste [MENU].
Page 47
– hPa – mmHg – mbar – psi – kPa Im Allgemeinen empfi ehlt METTLER TOLEDO die Verwendung von Absolutdruck- Im Allgemeinen empfiehlt METTLER TOLEDO die Verwendung von Absolutdruck- 20.9 %V O sensoren für eine genauere Signalkompensation über einen breiten Druckbe- 20.9...
Page 48
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C Ist die feste Kompensation ausgewählt, muss der für die Berechnung des Mess- Ist die feste Kompensation ausgewählt, muss der für die Berechnung des Mess- 20.9 %V O signals erforderliche feste Temperaturwert manuell eingegeben werden.
Quick Setup Guide TDL GPro™ 500 Einstellung Die Durchflussrate des Spülgases beeinflusst die effektive optische Weglänge der korrekten und damit den Messwert. Spülung der Daher ist die folgende Vorgehensweise unbedingt einzuhalten! Zu Beginn mit Prozessseite einer sehr hohen Durchflussrate anfangen und diese schrittweise verringern. Der Messwert ist zunächst sehr klein und steigt mit abnehmendem Spülgasfluss.
Quick Setup Guide TDL GPro™ 500 4 VerifizierungundWartung EinpunktkalibrierungvonTDL-Gassensoren Rufen Sie den Kalibrierungsmodus auf, siehe Abschnitt 7.1 «Justiermodus auf- 20.9 %V O rufen». 25.0 °C Justieren Sensor Eine Einpunktkalibrierung eines Gassensors ist stets eine Kalibrierung der Steil- Kanal B TDL heit (d. h.
Quick Setup Guide TDL GPro™ 500 Geben Sie den Konzentrationswert der Probe ein und drücken Sie dann die Taste 12.1 %V O [ENTER], um die Ergebnisse für die Kalibrierung zu berechnen. 25.0 °C B Punkt1 = 56.90 %sat = 57.1 %air Nach der Kalibrierung wird der Steilheitsfaktor «S»...
Quick Setup Guide TDL GPro™ 500 5 Fehlermeldungen Fehlermeldung Beschreibung Kein Sensor auf Kanal 3 M400 kann keinen bekannten ISM-Sensor feststellen. Wenn kein Sensor gefunden werden kann, erscheint die Meldung KEIN SENSOR ERKANNT im Display. Signalverarbeitung Störung Fit der Linienprofile fehlgeschlagen. Laser Störung Die Wellenlänge des Lasers hat sich verändert.
Page 53
– Wenn weiterhin nach 60 Sekunden die Zeitabschaltung er- Wenn weiterhin nach 60 Sekunden die Zeitabschaltung er- Wenn weiterhin nach 60 Sekunden die Zeitabschaltung er- folgt, senden Sie das Gerät zurück an METTLER TOLEDO. Senden Sie das Gerät zurück an METTLER TOLEDO Fehler...
Page 55
Quick Setup Guide TDL GPro™ 500 French GPro™ 500 Guide de paramétrage rapide Sommaire 1 Préparation Contenu de l‘emballage Exigences en ce qui concerne le site Conditions de fonctionnement ambiantes Ce dont vous avez également besoin Accessoires en option 2 Avant l‘installation Mise en place de la bride Exigences en ce qui concerne la bride Conditions de débit au point de mesure...
Quick Setup Guide TDL GPro™ 500 2 Avant l‘installation Mise en place de la bride 60 cm (23.6") La tête TDL doit être facilement accessible. Une personne doit pouvoir se tenir debout face à celle-ci et serrer les boulons de fixation M16 à l‘aide des deux clés standard.
Quick Setup Guide TDL GPro™ 500 DN100 DN50/PN25 DN65/PN25 100 mm 100 mm ANSI 4" ANSI 2"/300 lbs ANSI 2.5"/300 lbs 4" 4" Conditions de débit Lors de la détermination de l‘emplacement du GPro™ 500 TDL dans le Lors de la détermination de l‘emplacement du GPro™ 500 TDL dans le au point de mesure procédé, il est recommandé...
Quick Setup Guide TDL GPro™ 500 Purge 0…1 L /min 0…10 L /min Pressure regulator Rotameter 1 Arrivée du gaz de purge pour le côté instrument (raccord de tube de 6 mm ou de ¼"). 2 Arrivée du gaz de purge pour le côté procédé (clapet anti-retour requis). 3 Sortie du gaz de purge pour le côté...
Quick Setup Guide TDL GPro™ 500 Purge avec barrière thermique Configuration de la purge 1 Arrivée du gaz de purge pour la barrière thermique (raccord de tube de 6 mm ou de ¼") 2 Sortie du gaz de purge pour la barrière thermique (raccord de tube de 6 mm ou de ¼") 3 Arrivée du gaz de purge pour le côté...
Quick Setup Guide TDL GPro™ 500 Miseàlaterreet câ- blage (ATEX) Point de mise à la terre externe pour câbles > 4 mm (M6 × 12) Point de mise à la terre externe...
Quick Setup Guide TDL GPro™ 500 Entréesanalogiquesactives(versionATEX) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Entréesanalogiquesalimentéesparboucle(ATEX) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 Miseàlaterreetcâblage(FM) Point de mise à la terre externe pour câbles > 4 mm (M6 × 12) Point de mise à la terre externe Mise à la terre de protection Matériau : AISi7Mg0,3 chromé Taille : M6 × 12 mm 2 options de mise à...
Quick Setup Guide TDL GPro™ 500 Entréesanalogiquesactives(versionFM) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Entréesanalogiquesalimentéesparboucle(versionFM) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 CâblesGPro™500 Signal Description N°câble Couleur Boîte de Broche n° Broche n° jonction Alimentation Rouge + 24 V Alimentation 24 V, 5 W Terre Bleu (alimentation) RS 485 A Vert Interface M400 RS 485 B Jaune (RS 485) RS 485 GND...
Page 70
Quick Setup Guide TDL GPro™ 500 Pin 1 Pin 4 Connexions à la carte mère intégrée à la tête de l’électrode Remarque : Lecouvercledelatêtedel‘électrodeversionATEXnedoit jamaisêtreouvertcarcelaannulelacertificationATEX. Pin 1 Pin 12 Connexions à la carte E / S intégrées à la tête de sonde AVERTISSEMENT Toutes les ouvertures doivent être fermées avec des presse-étoupes ou des connecteurs de blocage certifiés répondant au même degré...
être détectées et acceptées. Si un paramètre spécial est choisi, il est le seul à être détecté et accepté sur le transmetteur. Étalonnage Le GPro 500 est étalonné en usine et n‘exige aucun nouvel étalonnage à l‘installation et au démarrage. InstallationdeTDL (Chemin d‘accès : Configuration rapide / TDL / Installation)
Page 73
– hPa – mmHg – mbar – psi – kPa En règle générale, METTLER TOLEDO recommande d‘utiliser des transducteurs 20.9 %V O à pression absolue pour une compensation des signaux plus précise sur une 20.9 %Trm large plage de pressions.
Page 74
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C Si une compensation fixe est sélectionnée, la valeur de température fixe avec 20.9 %V O laquelle le signal de mesure est calculé doit être saisie manuellement. Pour la 20.9 %Trm température fixe, seul le °C peut être utilisé.
Quick Setup Guide TDL GPro™ 500 Configurationde Le débit de la purge a une incidence sur la longueur de chemin effective, et par la purge correcte conséquent sur la valeur de mesure. côté procédé Il faut donc utiliser la procédure suivante. Commencer par un débit très élevé et le réduire progressivement.
Quick Setup Guide TDL GPro™ 500 4 Vérificationetmaintenance ÉtalonnageenunpointpourlessondesTDLpouroxygènegaz Accédez au mode d‘étalonnage comme indiqué dans la section 7.1 « Accès au 20.9 %V O mode Calibrage ». 25.0 °C Calibrer Capteur Pour les sondes à gaz, l‘étalonnage en un point est toujours effectué sur la pente Canal B TDL (autrement dit à...
Quick Setup Guide TDL GPro™ 500 Saisissez la valeur de concentration de l‘échantillon et appuyez sur la touche 12.1 %V O [ENTER] pour démarrer le calcul des résultats d‘étalonnage. 25.0 °C B Point1 = 56.90 %sat = 57.1 %air Après l‘étalonnage, la pente « S » est affichée. 12.1 %V O 25.0...
Quick Setup Guide TDL GPro™ 500 5 Messages d‘erreur Message Remarque Aucune sonde sur la voie 3 Le M400 n‘est pas en mesure de détecter les éventuelles sondes ISM qu‘il peut identifier. Si aucune sonde n‘est détectée, un message faisant état de l‘ABSENCE DE SONDE DÉTECTÉE apparaît.
Page 79
Procédez à un étalonnage avec le tube Anomalie Erreur logicielle correspondant. Si cela ne résout toujours pas le problème, retournez l‘unité à METTLER TOLEDO pour un remplacement de la carte flash. Vérifiez la sonde de pression externe et Demande de Erreur système...
Page 81
Quick Setup Guide TDL GPro™ 500 Portuguese GPro™ 500 GuiadeConfiguraçãoRápida Indice 1 Preparação Conteúdo da embalagem Requisitos de campo Condições ambientais de operação O que você também precisa Acessórios opcionais 2 Antesdainstalação Colocação do flange Requisitos do flange Condições de fluxo no ponto de medição Purga Purga com barreira térmica Aterramento e ...
Quick Setup Guide TDL GPro™ 500 2 Antesdainstalação Colocaçãodoflange 60 cm (23.6") A cabeça do TDL deve estar facilmente acessível. Deve ser possível que uma pessoa possa permanecer em frente dele e ajustar os parafusos de fixação M16, usando duas chaves de boca padrão. Deve haver, pelo menos, um espaço livre de 60 cm do flange fixado na tubulação e para fora, como mostrado a seguir.
Quick Setup Guide TDL GPro™ 500 DN100 DN50/PN25 DN65/PN25 100 mm 100 mm ANSI 4" ANSI 2"/300 lbs ANSI 2.5"/300 lbs 4" 4" Condiçõesdefluxo Ao decidir-se pela substituição do GPro™ 500 TDL no processo, é recomen- Ao decidir-se pela substituição do GPro™ 500 TDL no processo, é recomen- nopontodemedição dável uma distância mínima de 5 diâmetros em comprimento de duto reto antes e 3 diâmetros em comprimento de duto reto após o ponto de medida.
Quick Setup Guide TDL GPro™ 500 Purga 0 a 1 L /min 0 a 10 L /min Regulador de pressão Rotâmetro 1 Entrada do gás de purga no lado do instrumento (conexão de tubo de 6 mm ou ¼" ). 2 Entrada do gás de purga no lado do processo (deve ter uma válvula de retenção).
Quick Setup Guide TDL GPro™ 500 Purga com barreira térmica Configuraçãodapurga 1 Entrada do gás de purga da barreira térmica (conexão de tubo de 6 mm ou ¼" ) 2 Saída do gás da purga da barreira térmica (conexão de tubo de 6 mm ou ¼"...
Quick Setup Guide TDL GPro™ 500 Aterramentoe l igações elétricas ( ATEX) Ponto de aterramento externo para cabos > 4 mm² (M6 × 12) Ponto de aterramento externo.
Quick Setup Guide TDL GPro™ 500 Entradasanalógicasativas(versãoATEX) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Entradasanalógicasalimentadasporcircuito(ATEX) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 Entradasanalógicasativas(versãoFM) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 Entradasanalógicasalimentadasporcircuito(versãoFM) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 GPro™ 500 cabos Sinal Descrição Cabo n° Caixa de Pino nº Pino nº junção Alimentação Vermelho Alimentação em + 24 V em 24 V, 5 W GND (Energia) Azul RS 485 A Verde Interface M400 RS 485 B Amarelo (RS 485)
Page 96
Quick Setup Guide TDL GPro™ 500 Pin 1 Pin 4 Conexões na placa-mãe na cabeça do sensor. Nota: AtampadacabeçadosensordaversãoATEXnunca deveseraberta,poisistoinvalidaacertificaçãoATEX. Pin 1 Pin 12 Conexões na placa I / O na cabeça do sensor. ATENÇÃO Todas as aberturas precisam ser fechadas com prensa-cabos ou tam- pões certificados com o mesmo grau de certificação do GPro™...
Se um parâmetro especial for escolhido, somente esse parâmetro será reconhecido e aceito no transmissor. Calibração O GPro 500 é calibrado na fábrica e não requer calibração na instalação e inicialização. InstalaçãodoTDL (caminho: Quick Setup / TDL / Installation) No modo medição, pressione a tecla [MENU].
Page 99
Pressão=Externa – hPa – mmHg – mbar – psi – kPa Em geral, a METTLER TOLEDO recomenda o uso de transdutores de pressão 20.9 %V O absoluta para obter sinais de compensação mais precisos em uma ampla 20.9 %Trm faixa de pressão. Se, porém, forem esperadas pequenas variações na pressão 4 mA = 940.0 mbar...
Page 100
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C Se a compensação fixa for selecionada, o valor fixo da temperatura com que 20.9 %V O o sinal de medição será calculado deve ser digitado manualmente. Para a 20.9 %Trm temperatura fixa, somente pode ser usado °C.
Quick Setup Guide TDL GPro™ 500 Configuraçãocorreta A taxa de fl uxo de purga irá afetar o comprimento efetivo do caminho e, conse- A taxa de fluxo de purga irá afetar o comprimento efetivo do caminho e, conse- da purga no lado do quentemente, o valor da medição.
Quick Setup Guide TDL GPro™ 500 4 Verificaçãoemanutenção CalibraçãodeumpontoparasensoresTDLdegás Entre no modo de calibração conforme descrito na seção 7.1 «Acessar Modo 20.9 %V O de calibração». 25.0 °C Uma calibração de um-ponto dos sensores de gás é sempre uma calibração Calibrar Sensor Canal B TDL de slope (ou seja, com ar).
Quick Setup Guide TDL GPro™ 500 Insira o valor da concentração da amostra e pressione a tecla [ENTER] para 12.1 %V O iniciar o cálculo dos resultados da calibração. 25.0 °C B ponto1 = 56.90 %sat = 57.1 %air Após a calibração o slope “S” é exibido. 12.1 %V O 25.0...
Quick Setup Guide TDL GPro™ 500 5 Mensagens de erro Mensagem Comentário Nenhum sensor no canal 3 O M400 não é capaz de detectar sensores ISM que este pode identificar. Se nenhum sensor for encontrado ele exibirá a mensagem SENSOR NÃO DETECTADO Falha no sinal de processo Conexão dos perfis de linha falhou.
Page 105
Erro de software Realize uma calibração com o tubo de calibração. Se ainda assim não der certo, envie a unidade de volta para a METTLER TOLEDO para troca de Flashcard. Verifique o sensor de pressão externa Solicitação de Erro de sistema e o mapeamento manutenção...
Quick Setup Guide TDL GPro™ 500 有源模拟输入(ATEX 标准) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 回路供电模拟输入 (ATEX) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Quick Setup Guide TDL GPro™ 500 有源模拟输入(FM 标准) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND...
Quick Setup Guide TDL GPro™ 500 回路供电模拟输入(FM 标准) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A Sensor In Temperature sensor ...
Page 126
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C 如果选择了固定补偿,则必须手动输入计算测量信号所使用的固定温 20.9 %V O 度值。对于固定温度,只能使用 °C。 20.9 %Trm Temperature=Fixed 20.9 %V O 20.9 %Trm Temperature=320.0 °C 最后,选择与安装的探头长度相对应的初始光程长度: 20.9 %V O –...
Page 133
Quick Setup Guide TDL GPro™ 500 한국의 GPro™ 500 빠른 설정 가이드 만족 1 준비 포장 내용물 현장 요건 주변 작동 조건 그외 필요한 것 액세서리 옵션 2 설치 전 플랜지 설치 위치 플랜지 요건 측정 지점의 흐름 상태 퍼징 열...
Quick Setup Guide TDL GPro™ 500 1 준비 – GPro™ 500 TDL(파장 가변 다이오드 레이저) 분석기 포장 내용물 – 1 안전 지침 – 1 사용자 매뉴얼 및 MT-TDL Software Suite가 포함된 문서 CD – GPro™ 500 전원 공급 장치용 24 VDC, 5 W 현장...
Page 136
키트 - 평면 개스킷 ST 30 080 914 키트 - 평면 개스킷 HT(흑연) 30 080 915 케이블 - GPro 500 ATEX, FM 5 m 30 077 735 케이블 - GPro 500 ATEX, FM 15 m 30 077 736 케이블 - GPro 500 ATEX, FM 25 m...
Quick Setup Guide TDL GPro™ 500 2 설치 전 플랜지 설치 위치 60 cm (23.6") TDL 헤드는 쉽게 액세스 가능해야 합니다. 사람이 그 앞에 서서 2개의 표준 스패너로 M16 고정 볼트를 조정할 수 있어야 합니다. 아래와 같이 스택에 고정된 플랜지부터 바깥 쪽으로 최소 60 cm의 여유 공간이 측정되어야...
Quick Setup Guide TDL GPro™ 500 DN100 DN50/PN25 DN65/PN25 100 mm 100 mm ANSI 4" ANSI 2"/300 lbs ANSI 2.5"/300 lbs 4" 4" 공정에서 GPro™ 500 TDL의 설치 위치를 결정할 때 측정 지점 앞의 측정 지점의 흐름 거리는 직선 덕트 스택 지름의 최소 5배, 측정 지점 뒤의 거리는 직선 상태...
Page 139
Quick Setup Guide TDL GPro™ 500 퍼징 0…1 L/min 0…10 L/min 압력 조절기 로터미터 1 기기 측면의 퍼지 가스 입구(6 mm 또는 ¼" 튜브 피팅). 2 공정 측면의 퍼지 가스 입구(점검 밸브가 있어야 함). 3 기기 측면의 퍼지 가스 출구(6 mm 또는 ¼" 튜브 피팅). 4 공정...
Quick Setup Guide TDL GPro™ 500 열 경계부 퍼징 퍼징 구성 1 열 경계부에 대한 퍼지 가스 입구(6 mm 또는 ¼" 튜브 피팅) 2 열 경계부에 대한 퍼지 가스 출구(6 mm 또는 ¼" 튜브 피팅) 3 공정 측면의 퍼지 가스 입구(점검 밸브가 있어야 함) 4 필수...
Quick Setup Guide TDL GPro™ 500 활성 아날로그 입력(ATEX 버전) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B...
Quick Setup Guide TDL GPro™ 500 루프 전원 아날로그 입력(ATEX) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A...
Quick Setup Guide TDL GPro™ 500 접지 및 배선(FM) 4 mm 초과 케이블에 대한 외부 접지점 (M6 × 12) 외부 접지점. AISi7Mg0.3 : M6 × 12 mm : 1.4404(AISI 316L) : M4 × 6 mm 4 mm 보호 접지.
Quick Setup Guide TDL GPro™ 500 활성 아날로그 입력(FM 버전) User-provided ACTIVE ANALOG INPUTS Pressure – 100 mm (4") Temperature – – – RS-485 A Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B...
Quick Setup Guide TDL GPro™ 500 루프 전원 아날로그 입력(FM 버전) User-provided SENSOR IN Pressure 100 mm (4") Temperature SENSOR IN Ethernet Analog Outputs 24 VDC M400 24 VDC Power 0.2 A (5W) RS-485 A RS-485 GND RS-485 B M400 TB4 RS-485 B RS-485 GND RS-485 A...
Quick Setup Guide TDL GPro™ 500 GPro™ 500 케이블 신호 설명 케이블 번호 색상 정션 박스 핀 번호 핀 번호 전력 + 24 V 빨간색 전력 24 V, 5 W GND(전력) 파랑색 RS 485 A 녹색 인터페이스 M400 RS 485 B 노란색...
Page 148
Quick Setup Guide TDL GPro™ 500 Pin 1 Pin 4 센서 헤드 내 메인보드에 연결 참고: ATEX 버전의 센서 헤드 커버가 열리면 ATEX 인증이 무효화되기 때문에 절대 열려서는 안 됩니다. Pin 1 Pin 12 센서 헤드 내 IO 보드에 연결. 경고...
선택한 경우 ”Auto” 설정은 사용 가능한 모든 ISM 센서가 인식되고 수용 됩니다. 특수 파라미터를 선택한 경우 트랜스미터에 오직 이 파라미 터만 인식되고 수용됩니다. 교정 GPro 500은 공장에서 교정되어 출시되며 설치 및 시작 시 교정이 필요하지 않습니다. TDL 설치 (경로: Quick Setup / TDL / Installation) 측정...
Page 151
Pressure=External – hPa – mmHg – mbar – psi – kPa 일반적으로 METTLER TOLEDO는 폭넓은 압력 범위에 걸쳐 더욱 정확한 20.9 %V O 신호 보상을 위해 절대적으로 압력 트랜스듀서를 사용할 것을 권장합 20.9 %Trm 니다. 그러나 주변 대기압에 미미한 압력 변화가 예상되는 경우에는...
Page 152
Quick Setup Guide TDL GPro™ 500 20.9 %V O 20.9 %Trm 4 mA = 0.000 °C Ain 20 mA = 250_0 °C 고정 보상을 선택한 경우 측정 신호를 계산하는 고정 온도 값을 수동 20.9 %V O 으로 입력해야 합니다. 고정 온도의 경우 °C만 사용할 수 있습니다. 20.9 %Trm Temperature=Fixed...
Page 153
Quick Setup Guide TDL GPro™ 500 정확한 공정 측면 퍼 퍼징의 유속은 유효 경로 길이와 결과적으로 측정 값에 영향을 미치게 징 설정 됩니다. 그러므로 다음과 같은 절차를 사용해야 합니다. 매우 높은 유속에서 시작하여 점차 낮춥니다. 그리고 나면 낮은 값에서 측정 값이 시작되어 퍼지...
Quick Setup Guide TDL GPro™ 500 샘플의 농도 값을 입력한 다음 [ENTER] 키를 눌러 교정 결과의 계산을 시 12.1 %V O 작합니다. 25.0 °C B Point1 = 56.90 %sat = 57.1 %air 교정 후 기울기 ”S”가 표시됩니다. 12.1 %V O 25.0 성공적으로...
Quick Setup Guide TDL GPro™ 500 5 오류 메시지 메시지 설명 채널 3에 센서 없음 M400이 식별할 수 있는 ISM 센서를 감지할 수 없습니다. 센서를 찾지 못한 경우에는 감지된 센서 없음 메세지가 표시됩니다. 신호 처리 실패 라인 프로필 피팅 실패. 레이저...
Page 157
공정 내 먼지 부하를 줄입니다. 교정 튜브로 교정을 수행하십시오. 장애 소프트웨어 여전히 작동하지 않을 경우 플래시 카드 오류 교체를 위해 장비를 METTLER TOLEDO에 반납하십시오. 외부 압력 센서 및 맵핑 점검 유지보수 시스템 오류 요청 외부 온도 센서 및 매핑 점검...