Heine ® Ml4 Led With Dv1; English - HEINE ML4 LED Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 7
HEINE
ML4 LED with DV1
®
This manual is valid for the HEINE DV1 digital video camera in conjunction
with the HEINE headlight LED ML4.
Please read and follow these instructions for use of and keep them
for future reference.
Intended Use
The optional DV1 is a digital camera that provides via a USB 2.0 interface
contactless high resolution digital images and a fast live stream. It is used as
an educational tool and documentation to show the examiner view to a larger
audience or for recording medical techniques for teaching. The DV1 is not a
medical product and presents no images for diagnosis, pathological analysis,
operation or classification of health or disease.
For U.S. only:
Federal law restricts this device to sale by or on the order of a
Physician or Practitioner!
Warnings and Safety Information
Caution! Indicates potential hazardous situations. Ignoring the
corresponding instructions may lead to dangerous situations of mild
to moderate extent. (Background color yellow, foreground color black.)
Note! Note indicates valuable advice in terms of installation, operation,
maintenance or repair. Notes are important, but not related to hazardous
situations.
Product overview
1
2
3
1 Focus Lever
2 USB port (mini-USB)
3 Locking screw for camera tilt
4 Mounting for USB cable to the headband
5 Fastening USB cable to the back of the head band
(only for use with mPack)
Setting up
Only connect devices to the camera, which correspond at least to the
standard of IEC 60950-1 Information technology devices. And as well
have a UL listing or a GS mark.
Technical requirements
To ensure a smooth data flow for the video recording the connected
computers must meet the following minimum requirements:
- Processor: Intel Dual Core 2.2 GHz
- Graphics card: 24 bit
- Operating system: Windows XP, Windows 7
- Monitor: Resolution of 1280x1024 pixels or higher, refresh rate of at
least 60 hertz, colour depth, True Color (32 bit) colour calibrated
- DirectX 9.0c or higher
- USB 2.0 (when using a footswitch another USB port is required)
- USB 2.0 cable, length 3m (included)
Software Installation
Before first use, the installation of the software and the camera driver is
necessary. Please follow the installation instructions that come in the support
area (see section „LED ML4 DV1") of the website http://dv1.heine.com
4
5
Assembly
The following steps must be performed outside the patient environment.
Plug the USB cable into the mini-USB socket on the camera. Fix then the USB
cable to the head piece of the headband (4). In as far as you use the mPack
as a power source for the ML4 LED hang the USB cable, also behind the
head band (5). Put on ML4 LED with the DV1 and adjust the headband so that
it fits securely and comfortably. Switch on the light of the ML4 LED and loosen
the fastening screw for the camera tilt (3). Tilt camera so that the illumination
field is visible in the centre on the screen. Tighten again the locking screw for
the camera tilt.
Operation
The following steps must be performed or checked every time the
computer is restarted outside of the patient environment.
1. Start the computer
2. Connect the instrument to the power supply. Check the function by briefly
turning on the device.
3. Connect the USB cable.
4. Open IC Capture program on the computer.
5. The DV1 digital video camera is automatically detected by the computer.
Confirm your selection with „OK" .
6. Open configuration file.
7. Set video compression.
8. Put on the instrument and adjust headband for a comfortable fit.
9. Set DV1 digital camera and illumination field sequentially. For this purpose,
activate the light switch on the instrument.
Loosen locking screw for camera tilt (3). Tilt camera so that the illumination
field is visible in the centre on the screen. Tighten the locking screw for
camera tilt.
10. Adjust focus using the focus selector switch (1). Then look at an object that
is approximately in working distance. Possibly a second person is helpful
for this setting.
A detailed description of the video and photo recording incl. the specified
settings can be found in the installation instructions.
Video recording
1. Show toolbar or recording window
2. Start recording by clicking the Record button.
3. Stop recording with stop button. The data are stored automatically in the
predefined directory.
Photo Shoot
1. Activate the video live window by clicking
2. Press spacebar to store the video in the default folder.
Hygienic Reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national
standards, laws and guidelines.
Classification according to KRINKO: non-critical
Spaulding Classification USA: non-critical
Allow the device to cool down before reprocessing.
Before cleaning, disconnect the device from the power source.
In the event of suspected contamination, carry out hygienic preparation
of the instrument.
The device and its accessories are not sterile. Avoid touching with sterile
hands, areas or instruments.
The described cleaning and disinfection measures do not replace the
specific rules applicable for the establishment.
HEINE Optotechnik only approves the resources and procedures named
in these instructions for use.
Cleaning and disinfection may only be carried out by personnel with
sufficient hygienic knowledge.
Observe the instructions of the manufacturer of the reprocessing media.
Do not use any spray disinfectant, or any dripping-wet or heavily-foaming
cloths. Do not prepare mechanically.
med 2714
2014-12-15

ENGLISH

5/28

Publicité

Table des Matières
loading

Table des Matières