5.1.8.1 Connection pipes with water
While the tubes for condenser water are being connected, keep the following notes in mind:
-Use the connectors for red tubes that we provide with the equipment
-The tubes used must have a diameter equal to the connector, approximately 8mm, see section
Accessories and spare parts"
-For safety reasons, try to secure the tubes with clamps, we recommend that you look at section
Accessories and spare parts"
-Check the tubes from time to time and replace them when they are brittle or broken
5.1.8.2 Empty tubes
When connecting the tubes to perform the vacuum in the equipment, keep in mind the following
indications:
-Use the connectors for red tubes that we provide with the equipment
-The tubes used must have a diameter equal to the connector, approximately 8 mm, see section
Accessories and spare parts"
-Keep the vacuum tubes as short as possible
-The vacuum pipe does not need to be secured
-Check the tubes from time to time and replace them when they are brittle or broken
5.2 Operation
This chapter explains how to operate the equipment with the different elements and the possible modes
of operation properly and safely.
5.2.1 Motor height adjustment
For a more ergonomic working position, the height of the motor can be adapted to the user requirements.
If you want to modify the height, proceed as follows:
-Once the equipment is turned on, press the (up) or (down) button and the height will change
-If you want to finish the height adjustment, stop pressing the button and the movement will stop
automatically
5.2.2 Bath temperature adjustment
As soon as you turn on the device with the on / off button, the thermostatic bath will start to heat up if the
actual temperature is below the set temperature.
To adjust a temperature, press the "temp" button and, turning the control knob, select the needed
temperature. To confirm the action, press the control knob.
to see our references
to see our references
Warning
Risk of injury
Do not operate this instrument if the glass is
broken
Warning
The thermostatic bath can reach temperatures
above 100°C. To avoid burns, consider:
- To remove the evaporating flask from the bath
- To make sure that the liquid does not overflow
the bath when the bottle is submerged