Control panel
Refer to illustration B.
The control panel of your battery includes the following:
1.
Audible alert mute button
2. Charge level check button
3. DC on/off switch
4. DC input/output port
5. Audible alert mute indicator
Setting up
CAUTION
•
Ensure the RPS II and the device are turned off before setting up.
•
When the RPS II is plugged to the mains power supply, ensure all cables are
properly connected.
Charging the battery
Refer to illustration C.
1.
Attach the power supply unit adapter to the DC plug and connect into the battery
DC input/output port.
2. Connect the AC or DC power cord to the power supply unit.
3. Plug the other end of the AC or DC power cord into the mains power outlet.
Notes:
•
Charging takes less than 4 hours from 0% battery level to greater than 95%.
•
To discontinue charging, unplug the power cord from the mains power outlet.
Providing power to the device
Refer to illustration D.
1.
Select the correct output voltage for your device using the voltage output selector
from the RPS II rear panel.
2. Connect the appropriate DC output cable into the battery.
3. Connect the other end of the DC output cable into the device.
4. Set the DC on/off switch to on.
Providing backup power to the device (for systems with both AC and DC inputs)
WARNING
In this configuration, the Elisee draws power from the RPS II first until it is
fully discharged, and will not recharge. As a result, once all power has been
discharged from the battery, it will not act as an AC backup.
Refer to illustration E.
1.
Select the correct output voltage for your device using the voltage output selector
from the RPS II rear panel.
2. Connect the appropriate DC output cable into the battery.
3. Connect the other end of the DC output cable into the device.
4. Connect the AC power cord to the rear of the device.
2
6. Battery level indicators
7 .
Charging indicator
8. DC switch/discharging indicator
9. Voltage output selector switch (rear
panel)