English
The Max. value of SpO
the Min. value of SpO
Max. value of pulse rate, the
Min. value of pulse rate and the
memory interval mode (default
18
Default preview
value of 5 minutes) appear in
that order at an interval of 2.5
seconds. The device will then
exit the setup mode 2.5
seconds later.
Blood saturation & heart rate
appears (- -) and yellow
19
Unable to measure
backlight flickers once every
second.
7. Specifications
Performance
Scope of measurement
SpO
: 35% - 100%
2
Pulse rate: 30 - 250 bpm (beats per minute)
Precision
SpO
: 70% - 100%: ±2 digits (%)
2
35% - 69%: unspecified
Pulse rate: 30 – 250 ± 3 digits
Data Update period: Less than 2 seconds
Averaging: 4 for SpO
; 8 for Pulse rate
2
Alarm condition delay:
Less than 8 seconds for SpO
Less than 16 seconds for pulse rate.
Alarm signal generation delay:
Less than 1 seconds for SpO
Probe type
Probe: Solaris S100A
Extension cord: Solaris AC031-113
Electrical specifications
Battery (4 "AA" Alkaline batteries)
Battery capability: Can be used for 15 hours continuously depending on the type of
battery used.
Environmental conditions
Operating temperature: 5℃ - 40℃ (41℉ – 104℉)
Storage temperature:
Packaged: -20℃ - 70℃ (-4℉ – 158℉)
Naked: -20℃ – 60℃ (-4℉ – 140℉)
Relative humidity: 15% - 95% (no condensing)
, then
2
, the
2
None
Press
Beep-beep sounded
None
repeatedly.
.
2
and pulse rate.
2
15