Riester ri-vital Mode D'emploi page 80

Masquer les pouces Voir aussi pour ri-vital:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Set language
Set time
Set date
Delete memory
Leakage test
Manometer mode
Temperature unit
Time format
Keep pressing the menu button until Start is highlighted yellow in the manometer
mode. After pressing the Setup button the device will be in manometer/calibration
mode for 170 s.
b) Conduct the test in pressure ranges of at most 50 mmHg between 0 mmHg and the
highest value of the display range.
Exceeding of the overpressure can take place at any desired level. The test can be
continued after decreasing the overpressure to the desired value.
Press the On/Off button to exit the menu.
1. Control manometer (see 1.1b)
2. Blood pressure device to be tested
3. Metal container (see 1.1a)
4. Pump ball with discharge valve (see 1.1c)
Fig. 1: Test equipment to determine the error thresholds of the reading display of the
cuff pressure.
ri-vital
TECHNICAL MEASUREMENT CHECK
80
Only for technical service
Only for technical service

Publicité

Table des Matières
loading

Table des Matières