7.4 Pulse display
If the patient is connected to the pulsoximeter, the pulse reading that is obtained in this
manner is always indicated.
If no SpO2 measurement takes place, then after a successful blood pressure
measurement the device shows the oscillometrically determined pulse. Optionally,
the
ri-vital
amplification and artefact suppression).
8. BODY TEMPERATURE MEASUREMENT WITH THE

PREDICTIVE THERMOMETER

8.1 Probe covers – application and removal
A)
Open the probe cover box by lifting and pulling the tab on the upper corner.
B)
Insert the box with the probe covers from above into the slot. (In order to avoid
infections, never interchange or use opened boxes that have been used with blue or
red thermometer probes).
C)
Remove the thermometer probe from the slide compartment. This automatically
switches on the thermometer.
D)
As a reminder to apply or remove the probe cover, a probe symbol with flashing
probe cover is displayed before applying or removing the probe cover.
.
E)
Insert the probe tip of the thermometer probe into a probe cover in the box. Press
the grip of the thermometer probe until the probe cover noticeably clicks into place.
F)
Take the relevant temperature reading (oral, axillar or rectal temperature).Please
use the respective thermometer probe for the purpose.
The measurement is concluded once the final temperature value is displayed
G)
Press the upper button on the thermometer probe to eject the used probe cover into
a waste container.
H)
Remove, dispose of and replace the empty box.
ri-vital
PREDICTIVE THERMOMETER
60
can be ordered with Pearl pulsoxymeter technology (a process for pulse

Publicité