Select the desired option with
3
with
· On: The flow measurement value is displayed in the
lower measurement value line
· OFF: The flow measurement value is not displayed
in the lower measurement value line
OFF is selected:
Next handling objective
On is selected:
- The set unit of flow blinks.
Select the desired unit with
4
with
- The set air-density and the corresponding unit are
displayed.
Entering the air-density is required to calculate the flow
correctly.
Set the value with
5
- The set Pitot factor is displayed, Pitot factor lights
up.
The Pitot factor is dependant on the Pitot tube used.
Set the value with
6
4
Setting damping:
When damping is activated, a moving mean value
is shown in the display, which is created from an
adjustable number of measurement values:
· · 1 = Damping deactivated, current measurement
value is shown
· · 20 = Maximum damping, mean value calculation
over the last 20 measurement values.
The configuration modus is open, Damping lights up.
Set the value with
info@Testo-Direct.com
:
:
/
/
/
www.
/
.
4 S
D
ET
AMPING
/
and confirm
and confirm with
and confirm with
and confirm with
Testo-Direct
5. Operation 31