SI Analytics TitroLine 7500 KF Mode D'emploi page 136

Titrateur
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

136
Command
Description
aaAA
automatic allocation of device address
aaMC1...XX
choosing a method
aaBF
"filling burette". Aufsatz wird gefüllt.
aaBV
output of dosed volume in ml
aaDA
dose volume without filling, with adding the volume
aaDB
dose volume without filling, reset of the volume
aaDO
dose volume with filling, without adding the volume
aaGDM
dosing speed in ml/min
aaGF
filling time in seconds (min is 20, default 30)
aaES
"ESC" function one step backwards
aaEX
"exit" function.back to main menu
aafd
µa "dead stop" measurement function
aagdm
dosing speed in ml/min (0.01 – 100 ml/min)
aaGF
filling time in sec (adjustable 20 – 999 seconds)
aaGS
output serial no. Of device
aaLD
output of the measurement data
aaLR
output report (short report)
aaM
output the measurement value µA
aaLI
output method content
aaLO
output documentation (as configured)
aaRH
request of identification
aaRC
send last command
aaRS
report status
possible answers are:
"STATUS:READY" for ready
"STATUS:dosing" dosing
"STATUS:filling" filling
"ERROR:busy" if no interchangeable unit has been attached
aaSM
start selected method
aaSEEPROM
EEPROM reset to factory defaults
aaSR
stop the actual function
aaSS
titration start with the transfer of the pH end value
aaSYS5
adjust language to "German"
aaSYS1
adjust language to "English"
aaSYS2
adjust language to "French"
aaSYS3
adjust language to "Spanish"
aaVE
Version number of the software
Chapter 6 Data Communication via RS232 and USB-B interface
Reply
aaY
aaY
aaY
aa0.200
aaY
aaY
aaY
aaY
aaY
aaY
aay
aay
aaY
aaY
aaGS08154711
aaY
aaY
aaM0.1000
aaIdent:TL7500KF
aa"last command"
aaStatus:"text
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaVersion

Publicité

Table des Matières
loading

Table des Matières