Chapter 4 Method Parameters
125
Fig. 100
If"off"is selected for filling, filling it will not occur automatically after each dosing step.
If "intelligent before" is selected for filling, a verification will be performed each time prior to the next dosing step
in order to determine whether the dosing step can still be made without a filling operation. Should this prove to be
impossible, the first thing to occur is filling, followed by the dosing step.
If "intelligent after" is selected for filling, a verification will be performed after the next dosing step to find out
whether the next dosing step can still be made without filling.
If „always" is selected for filling, filling will occur automatically after each dosing step.
4.6.6

Sample identification

In the manual titration and in the preparation of solutions it is possible to input a sample identification. The
possible input includes manual, automatic or no sample description at all.
Fig. 101
For a sample description of the 'manual', a prompt for the sample description will always be displayed at the start
of the method (Cp. also chapter 3.6, Main menu). For an 'automatic' sample description there will be selected a
master description (in the current case this is water, cp. Fig. 102), which will then automatically be numbered
starting on 01.
Fig. 102
After a new power-up, numbering will resume with 01.