122
Fig. 94
The values can be set between 40 and 220 m. The pre-setting is 100 mV.
Low polarisation voltage:
High polarisation voltage:
4.6.4 Titration end
The end of a titration is reached, and the result will be calculated as soon as, or if, respectively:
•
The defined End value in µA value has been reached
•
The Endpoint delay in seconds has been adhered
•
The drift value in µg/min has been reached
•
The predefined value ml has been reached (Maximum titration volume)
•
The conditions for minimum and maximum titration time are maintained
Fig. 95
Fig. 96
Maximum titration time
The maximum titration time can be set between 0 – 9999 seconds. The default setting is 600 seconds. The
maximum titration time is generally used for KF titration, which can create a high continuous drift from a
secondary reaction and thus cannot reach a stable endpoint.
Chapter 4 Method Parameters
insensitive
sensitive