WARNING
Do not try to perform therapy without the super vision of your
healthcare professional.
WARNING
NPWT must run 24 hours a day. Do not stop the therapy
unless instructed by your healthcare professional or if exces-
sive bleeding occurs.
WARNING
The dressing should be changed and therapy restarted by
the healthcare professional if therapy is discontinued for
more than 2 hours.
WARNING
Invia Liberty pump does not detect blockage along tubing
connected to the drain adapter or connected to the curved
arm of the Y-connector. Check dressing tubing connected to
the drain adapter or the curved arm of the Y-connector fre-
quently throughout the day to make sure tubing is not
kinked or blocked.
If blockage is identified along tubing connected to the drain adap-
ter or tubing connected to the curved arm of the Y-connector,
contact your healthcare professional.
Turn OFF
Switch Invia Liberty pump OFF by pressing
[
] for 3 seconds. Press only once.