5) Allow to cure for at least 4 hours.
INFORMATION: Final bonding strength will be reached after
16 hours.
Improving corrosion protection
Polish or sinter the product for improved corrosion protection. Ottobock
recommends the 618T40* sintering powder for surface coating.
► Do not heat the material longer than 5 minutes at max. 150 °C for sur
face coating.
Using protective devices
► Use joint protectors to protect the product against soiling.
6 Cleaning
Promptly clean the product after contact with water containing salt, chlorine
or soap, or if it gets dirty.
1) Rinse the product with clear fresh water.
2) Dry the product with a cloth or allow it to air dry. Avoid exposure to direct
heat (e. g. from an oven or radiator).
7 Maintenance
INFORMATION
The product may be exposed to increased loads by the patient.
► Shorten the maintenance intervals according to the expected loads.
The manufacturer requires at least a semi-annual inspection of the product
to verify functionality and check for wear.
Spare parts are listed in the section "Components/design".
Only use 633F7 special lubricant.
Replacing the ball bearing
NOTICE
Canting the ball bearing
Mechanical damage
► Do not cant or damage the ball bearing while inserting it into the orthot
ic joint.
1) Loosen the screws and disassemble the orthotic joint.
2) Remove the ball bearing from the orthotic joint and replace it with a new
one.
3) Assemble the orthotic joint and tighten the screws.
13