1 Technical data
KERN (Type)
Trademark
Readability (d)
Weighing range (max)
Reproducibility
Linearity
Smallest part weight for
piece counting
Reference quantities at
piece counting
Weighing Units
Recommended adjustment
weight, not added (class)
Warm-up time
Stabilization time (typical)
Operating temperature
Humidity of air
Housing (B x D x H) mm
Dimensions of wind screen
(B x D x H) mm
Weighing plate,
stainless steel (mm)
Weight kg (net)
Electric Supply
Interfaces
4
PFB 120-3A
PFB 120-3
0,001 g
120 g
0,001 g
±0,003 g
2 mg
10, 20, 50, 100, 200
g, ct, lb, oz, d, ozt, dwt, mo, tl h, tl c, tl t, t, bt, n
100 g (F1)
+ 15° C .... + 35° C
max. 80 % (not condensing)
315 x 210 x 156
124 x 119 x 80
Ø 80
Input voltage weighing scales 12 V / 500 mA
Input voltage power supply unit 100 V – 240 V, 50/60 Hz
Bluetooth 2.0 (factory option)
Bluetooth 4.0 (factory option)
PFB 200-3A
PFB 300-3A
PFB 200-3
PFB 300-3
0,001 g
200 g
0,002 g
±0,005 g
±0,005 g
2 mg
200 g (F1)
300 g (F1)
2 hours
3 sec.
Ø 80
2 kg
RS232
PFB-BA-e-1741
0,001 g
300 g
0,002g
2 mg
Ø 80