• have an abrasive effect (such as sand).
Using this medical device under such conditions will render all claims
against Otto Bock HealthCare null and void.
The material is not UV or hydrolysis resistant.
The surface of the foot must not be damaged by grinding as this leads
to premature wear.
Please forward this information to your patients.
3 Description and functional wualities
A specially designed rollover element together with a sophisticated polyu-
rethane foam provide the main functions of this foot. The 1D10 is designed
for use with modular prostheses and is delivered with an integrated titanium
modular adapter.
The rollover element and foam technology offer the following functional
advantages:
• Comfortable heel strike
• Elastic return of the forefoot thanks to sophisticated foam technology
• Rollover smoother than with SACH foot due to flixible keel design
4 Handling
4.1 Alignment recommendations
4.1.1 Basic alignment recommendation
Foot size
Mid-foot anterior to alignment line
22-30 cm
30 mm
4.1.2 Static alignment recommendations
Align the prosthesis using the LASAR Posture, if available.
Follow the Ottobock alignment recommendations when aligning lower limb
prostheses.
4.1.3 Dynamic alignment recommendations
Align the prosthesis in the coronal (ML) and sagittal (AP) plane through slid-
ing or by changing the angle to achieve proper heel contact, smooth rollover
Heel height
10 mm +/- 5 mm
Ottobock | 7