Fig.
Item
–
–
–
–
–
–
–
–
–
–
5 Preparation for Use
CAUTION
Incorrect alignment, assembly or adjustment
Injuries due to incorrectly installed or adjusted as well as damaged prosthetic components
►
Observe the alignment, assembly and adjustment instructions.
5.1 Preparation
1) Turn on the prosthetic foot by pressing the main switch (see Page 26).
2) INFORMATION: The LED only indicates the charging process if the prosthetic foot is
switched on.
Charge the battery of the prosthetic foot (see Page 25).
3) Pull the footshell over the prosthetic foot (see Page 30).
5.2 Alignment
INFORMATION
To perform the prosthesis setup, the Galileo™ App is required.
►
Establish the Bluetooth connection between the Galileo™ app and the prosthetic foot as
described in the instructions for use of the Galileo™ app. You can find the instructions for
use in the main menu of the app under the Help menu item.
Activate the mode for orthopaedic technicians with the activation code on the sticker in this
►
document. (Settings>Activate Prosthetist Mode).
If the user would like to control the prosthetic foot with the Galileo™ app: delete the
►
user's Bluetooth coupling to activate the memory (see Page 28).
5.2.1 Bench Alignment
TT bench alignment
Required tools and materials:
662M4 goniometer, 743S12 heel height measuring apparatus, 743A80 50:50 gauge,
alignment apparatus (e.g. 743L200 L.A.S.A.R. assembly or PROS.A. 743A200
assembly)
Perform the alignment and assembly of the prosthetic components in the alignment apparatus
according to the following specifications:
Sagittal plane
Heel height:
Effective heel height (shoe heel height – sole thickness of forefoot) + 5 mm
1C66 Triton smart ankle
Quantity
Designation
1 ( )
EU AC adapter
US AC adapter
1 ( )
Power supply
1 ( )
Mini USB cable
1 ( )
Ankle cover
1 ( )
Ankle cover cap
Bench alignment process
Reference number
757S7*
757L2
625W20
2Z504
2Z505
Ottobock | 23