When changing the electrolyte, the membrane body
or the interior body, please observe the following
i nstructions (see also the following illustration):
a
Attention! Make sure that this maintenance step is
carried out in clean place.
1. Unscrew the cap sleeve from the sensor shaft and
carefully pull it off the sensor.
2. Pull off the membrane body from the interior
body. If it is tight-fitted, eject by pushing it with
the flat fi nger tip. Before electrolyte is refilled, the
m embrane body must be removed from the cap
sleeve!
3. Rinse the interior body with demineralized water
and carefully dab it dry with a paper tissue.
h
Note: steps 4 to 7 may only be carried out when
changing the interior body.
4. Unscrew the retainer nut of the interior body with
an adjustable wrench or with a 9 mm wrench.
5. Remove the interior body by pulling it out of the
sensor shaft. If necessary use a plier.
a
Warning! Do not twist the interior body. Other wise
the connection pins can be damaged.
6. Insert the new interior body in the sensor shaft.
Turn the interior body in the shaft until the slit of
the i nterior body is aligned with the pin placed in
the shaft.
7. Press the body in the shaft and screw the new
r etainer nut in place.
8. Examine the O-rings visually for mechanical
d efects, and replace if necessary.
9. Half-fill the new membrane body with O
e lectrolyte.
h
Note: The electrolyte bottle is equipped with
a special pouring system. To ensure proper
functioning, hold the bottle vertically, upside-
down.