Removing the data interface
– It is absolutely necessary to first disconnect the power cable from the data interface housing.
– Carefully swing housing straight back and down.
– Disconnect the ribbon cable connector (7).
– Unhook the tabs (6) from the balance rear wall.
– Cover the ribbon cable socket on the balance rear wall with the cover.
Adusting the power supply voltage
Note: The power supply voltage must be set on the housing of the data
interface.
The voltage selector switch can be moved to the proper position by using a
screwdriver:
Switch position
115 V
220 V
Installing the data interface housing
– Disconnect the power cable from the balance.
– Insert housing with the tabs (6) into the openings in the rear wall of the
balance.
– Plug in the ribbon cable connector (7). The ribbon cable must lead in
from the bottom.
– Move the top of the housing towards the balance until the power
connector (8) is completely plugged in.
– Reconnect the balance with the data interface to the power supply: plug
power cable into socket (9).
Leerseite
Connecting the instruments
The data interfaces of the 012 Option can be used simultaneously. If this is
done, please be aware that the configuration (baud rate, parity, operating
mode) is the same for both data interfaces. The input currently unused must
be in the BREAK mode to prevent the other interface from being blocked
(RS232C: space, CL: no current).
– Instruments with 20 mA current loop data interfacing are connected to
socket (11).
– Instruments with RS232C data interfacing are connected to socket (12).
(Non-METTLER TOLEDO instrument is DTE).
– If the data transfer is to be triggered by hand ("send on transfer" mode),
handkey (42500) or foot pedal (46278) is connected to sockets (10).
Voltage range
92...132 V
184...265 V
100 mA slow-blowing microfuse
2