Permanent cuff pressure, e.g. due to a bent cuff tube, may lead to
circulatory disorders that may cause serious injury to the patient.
If the cuff pressure remains, open the cuff fastener and switch the
device off by pressing the
Press the
- or the
process. The EN200 BP immediately stops the inflation process,
starts to discharge air.
To manually deflate the cuff, pull the tube off at the device's connector.
Makes sure that the tube is not bent, so that the cuff can be fully deflated.
The cuff must not be applied to wounds or cuts as this involves a higher
risk of infection.
Please note! The positioning and inflation of the cuff on any limb for
which an intravascular access or intravascular therapy or an arterio-
venous (a.v.) shunt is present may lead to a temporary interruption
and patient injury.
Please note! The inflation of the cuff may lead to a temporary loss of
function of a monitoring ME device used simultaneously on the same
limb.
Please note! The cuff tube may wrap around the neck and result in
strangulation. Please only wear the cuff on the upper arm and ensure
that the pressure tube cannot wrap around the neck.
Please note! Check that the operation of the automatic, non-invasive
blood pressure monitor does not impair the patient's blood circulation
for an extended period of time. (The check may include the observation
of the relevant limb.)
Please be aware that the method of measurement involves the
interruption of blood circulation for a short period. A break of at least
one minute must be provided between multiple measurements.
Measurements performed too frequently may injure the patient as
a result of circulatory disorders.
The results of the blood pressure measurement may also be influenced
by extreme temperatures and humidity. Pay attention to the operating
conditions.
If multiple unexpected measurement values occur, please measure
with an alternative device.
The device is only intended for wall mounting and may only be used
once it is mounted.
-button.
-button in order to terminate the inflation
MED 113275
2018-02-26
59/212