Faro MC01 S Mode D'emploi page 52

Piece a main
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MC01 S - Instructions for use - EN
CAUTION
Contact corrosion due to moisture.
Damage to product.
 Immediately remove the product from the steam steriliser after the sterilisation cycle!
The Faro medical device has a maximum temperature resistance up to138° (280.4 °F).
(Depending on the available autoclave,) select a suitable procedure from the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravitation method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F); in alternativa
– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
 Use according to the manufacturer's Instructions for Use.
6.8 Storage
Prepared products must be stored, protected from germs (as far as possible) and dust, in a dry, dark, cool room.
 Comply with the expiry date of the sterilised items!
50

Publicité

Table des Matières
loading

Table des Matières