MC01 S - Instructions for use - EN
Attach medical device to the motor coupling and turn it until the latch audibly snaps into place.
Pull on the medical device to make sure that it is securely affixed to the coupling.
5.2 Remove the medical device
Unlock the medical device from the motor coupling by twisting it slightly and then pulling it along its axis.
5.3 Insert the handpiece or contra-angle handpiece drill bit
NOTE
Only use handpiece or contra-angle handpiece drill bits that correspond to DIN EN ISO 1797-1, are made of steel or hard metal, and meet the following
criteria:
- Shaft diameter: 2.334 to 2.35 mm
with bit stop:
- Shaft clamping length: 12 mm
- Shaft clamping length: max. 22 mm
without bit stop:
- Shaft clamping length: min. 30 mm
- Shaft clamping length: max. 44.5 mm