8. Ensure clamp on each tube is open (Fig. 27) and position clamps away from patient�
9. Turn on power to the V�A�C�Ulta™ Therapy Unit, select the prescribed therapy settings, and initiate
therapy� Refer to the V�A�C�Ulta™ Therapy System User Manual for detailed instructions�
NOTE: The Test Cycle tool on V.A.C.Ulta™ Therapy unit may be used to confirm that the system was
set up correctly. Refer to the V.A.C.Ulta™ Therapy System User Manual for instructions on how to set up
therapy and use the Test Cycle tool.
10. The V�A�C� VeraFlo™ Dressing should have a wrinkled appearance shortly after therapy is initiated�
There should be no hissing sounds� If there is any evidence of leaks, check the seals around the
V�A�C� VeraT�R�A�C�™ Pad, or V�A�C� VeraT�R�A�C� Duo™ Tube Set pads, and V�A�C�® Advanced Drape,
tubing connections, canister connections, V�A�C� VeraLink™ Cassette connections, and ensure all
tubing clamps are open�
NOTE: The Seal Check™ tool on the V.A.C.Ulta™ Therapy Unit may be used to check for leaks in the
system. Refer to the V.A.C.Ulta™ Therapy System User Manual for instructions on how to use the Seal
Check™ tool.
NOTE: If a leak source is identified, patch with additional V.A.C.® Advanced Drape to ensure seal integrity.
11. Secure excess tubing to prevent interference with patient mobility�
NOTE: If the wound is over a bony prominence or in areas where weight bearing may exert additional
pressure or stress to the underlying tissues, a pressure-relief surface or device should be used to optimize
patient offloading.
25