Electromagnetic Compatibility; Guidance And Manufacturer's Declaration - MASCOT 205240 Mode D'emploi

De batteries lifepo4
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
EN

Electromagnetic Compatibility

In order to regulate the requirements for EMC (electromagnetic compatibility) with the aim of preventing
unsafe product situations, the EMC EN60601-1-2 standard has been implemented. This standard defines
the levels of immunity to electromagnetic interference as well as maximum levels of electromagnetic
emissions for medical devices. Medical devices manufactured by Mascot have been tested and conform
with the requirements of IEC/EN 60601-1-2, 3rd & 4th edition, nevertheless, special precautions may
need to be observed:
The Mascot products are suitable for use in Domestic, Residential, Office and Hospital environments,
except in special locations where EM Disturbances are known to be high, such as near High Frequency
Surgical Equipment or Magnetic Resonance Imaging systems.
When used according to its specification the User can expect the product to fulfil its essential performance,
being powering Medical Electrical Devices or charging batteries for Medical Electrical Devices.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoi-
ded because it may result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than provided by the manufactu-
rer could result in increased electromagnetic emissions or decreased electromagnetic im-
munity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the power supply or battery charger, including cables. Otherwise, degra-
dation of the performance of this equipment could result.

GUIDANCE AND MANUFACTURER'S DECLARATION

The Masot products are intended for use in the electromagnetic environment specified below. The custo-
mer or user should ensure that it is used in such an environment.
Test / Standard
Emission:
RF emissions,
CISPR 11
Harmonic emissions,
IEC 61000-3-2
Voltage fluctuations /
flicker emissions,
IEC 61000-3-3
Compliance level
Group 1, Class B
-
-
12
Guidance
Suitable for use in all establishments, inclu-
ding domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buil-
dings used for domestic purposes.
RF emissions not likely to cause any interfe-
rence in nearby electronic equipment.
However, a separation distance of 30 cm
shall be maintained.

Publicité

Table des Matières
loading

Table des Matières