Balluff BMD 1H Serie Notice D'utilisation page 110

Masquer les pouces Voir aussi pour BMD 1H Serie:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6 Set up with the display and adjustment module
Setup - Application - Gas
phase
Setup - Application - Di-
electric constant
Setup - Max. adjustment
Level
28
You can choose between the following medium types:
Dielectric con-
Type of medium
stant
> 10
Water-based liq-
uids
3 ... 10
Chemical mix-
tures
< 3
Hydrocarbons
This menu item is only available, if you have chosen interface meas-
urement under the menu item "Application". In this menu item you can
enter if there is a superimposed gas phase in your application.
Only set the function to "Yes", if the gas phase is permanently pre-
sent.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". In this menu item you can
enter the dielectric constant of the upper medium.
You can directly enter the dielectric constant of the upper medium or
have the value determined by the instrument.
If you want the dielectric constant to be determined by the instrument,
you have to enter the measured or known distance to the interface.
In this menu item you can enter the max. adjustment for the level. With
interface measurement this is the maximum total level.
Adjust the requested percentage value with [+] and store with [OK].
Examples
Acids, alcalis, water
isocyanate, chloroform
Solvents, oils, liquid gas

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Bmd 1l serie

Table des Matières