FloTech Programmable CheckMate Retain / Overfill Monitor
All active overfill sensors are dry.
All inactive overfill sensors are not connected.
All active retain sensors are dry or the Timer/Reset button
has been pressed.
All inactive retain sensors are disconnected.
The AUX input is jumpered or made through an external
contact.
If ANY of the above conditions are not met, the PERMIT signal
will not be generated.
Programming is accomplished in the following manner:
Procedure
To program the FT208 monitor:
1. Turn the power to the unit off.
2. Remove the reset switch wires from TB2.
3. Connect the programming module (10459) into TB2
bottom four contacts marked "Program, GND, Timer,
GND". Switches should be face up.
4. Move SW1 to "PRGM" position.
5. Turn on the power. The sensor LEDs will sequentially
flash through the power up test pattern. This lights the
sensor LEDs individually from TOP #1 through TOP #8
and then through BOTT #1 through BOTT #8.
6. The sensor LEDs will light up the last valid configuration.
If the memory does not contain a valid configuration, the
sensor LEDs will default to lighting the top 1 sensor.
7. Press SW2 pushbutton to light the RED Top and Bottom
Sensors light to match the total number of top and bottom
sensors. Each time SW2 is pressed, the sensor LEDs will
shift to the next valid configuration. This sequence is Top
#1 to Top #8 and then Top and Bottom #1 to Top and
Bottom #8. This sequence will repeat every time the
pushbutton is pressed.