NSK SGS Serie Mode D'emploi page 24

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CAUTION
• Clean and lubricate the handpiece prior to sterilization. If blood remains on the internal surface it can become clotted and cause product failure.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• Be sure to use sterilizers that can perform sterilization up to 135˚C. In some sterilizers, the chamber temperature may exceed 135˚C. Do not use these
sterilizers as failure of the handpiece could occur. Contact the sterilizer manufacturer for detailed information about cycle temperatures.
• Steam sterilization is recommended for the product. The validity of other sterilization methods (such as plasma sterilization or EOG sterilization) is not
confirmed.
9-5 Storage
Store the product in a dry, clean location.
CAUTION
• After the sterization and drying cycles are complete, remove the handpiece immediately from the sterilizer to store it.
• Store the product in a well ventilated place out of direct sunlight and within the range of temperature, humidity and pressure specified in "11. Specifications".
• Sterilization is not guaranteed after the sterilization retention period specified by the manufacturer and seller of the sterilization pouch has elapsed. If the
sterilization retention period has elapsed, perform sterilization again with a new sterilization pouch.
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sga serieSgs-esSgs-e2sSga-esSga-e2s

Table des Matières