6) Remove the handpiece from the motor. (Fig. 10)
(Refer to "4-2 Disconnecting")
7) Carry the handpiece and the irrigation nozzle to the decontamination area.
9-2 Cleaning, Disinfecting, and Drying
Perform cleaning, disinfecting, and drying using either the manual or automatic method.
Manual Method (Cleaning, Disinfecting, and Drying the Exterior)
Wash the handpiece and irrigation nozzle under running water following the procedures below.
(Water condition: ≤ 38˚C, ≥ 3.5L / min; water should be of the same quality as drinking water)
Fig. 10
15