Automatic Method (Cleaning, Disinfecting, and Drying the Exterior and Interior)
NSK handpieces with this symbol are compatible with medical thermo-disinfectors.
Due to the variation in cleaning / disinfection cycles and cleaning agents available from the various manufacturers, NSK has validated the Getinge 46 Series Washer
Disinfector Cycle P7-Anesthesia for automatic processing / reprocessing of the product.
Only validated procedures must be used for cleaning, disinfection and sterilization. When using reprocessing procedures that are different from those described in this
manual, those procedures must be validated by the respective practice or hospital using a thermo-disinfector that conforms to ISO 15883-1.
NOTICE
• All water is "building supply" (at least drinking water quality) unless otherwise indicated.
1) Prepare and pre-disinfect the device as described above in step "9-1 Cleaning at point-of use".
2) Wipe off a dirt on a part held by a handpiece holder using the cloth moistened with 60 - 90% ethanol or disinfectant.
When using disinfectant, follow the instructions given by the manufacturer of the disinfectant.
Use a state-sanctioned disinfectant with proven bactericidal, fungicidal, and virucidal properties.
The following disinfectants can be used in the United States and Canada.
®
®
CaviCide
, CaviWipes
(manufactured by Metrex)