NOTICE
• NSK recommends the use of "Spray Mist Absorber" (Y900084) to prevent oil mist expelling out of the handpiece head.
9-4 Packaging, Sterilizing, and Drying
1) Insert the handpiece into an FDA-cleared sterilization pouch that conforms to ISO 11607-1 and seal the pouch.
2) After packaging in a sterile pouch, sterilize using the following validated cycles. Use an FDA-cleared steam sterilizer to perform sterilization.
Follow local rules, regulations, and guidelines regarding the reprocessing of devices.
Type
Temperature
Full Cycle Time
Drying Time
CAUTION
• Use an FDA-cleared steam sterilizer to perform sterilization.
• Do not touch the product immediately after steam sterilization as it will be very hot and must remain in a sterile condition.
• Do not perform steam sterilization on the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to
the product from chemical residue on other instruments.
22
Gravity Displacement
Pre-Vacuum (Dynamic Air Removal)
132˚C
15 min.
30 min.
132˚C
4 min.
30 min.