11.2 Installation or replacement of the resin cartridges [3] and [4]
The duration of the resin cartridges depends on the quality of the feed water and the consumption of purified water.
Replace the cartridges when the warning-Replace the deionization resin is activated. That is, when the following
symbols are activated:
IMPORTANT! Before replacing the cartridges, verify the compliance of the Consumable storage requirements
described in Section 4.4 in the cartridges to be installed or call your supplier for the acquisition of new cartridges.
IMPORTANT! To install or replace the cartridges, ensure no water is being fed into the equipment and the
equipment is depressurized (follow instructions below). The QC connectors cannot be disconnected under pressure.
Follow the instructions below to install or replace the resin cartridges:
Close the feed water shut-off valve.
The warning Water cut is activated. Pressing the MODE / ESC key deactivates the acoustic signal.
Access the location of the resin cartridges (Cartridges [3] and [4]): The cartridges are installed inside the
equipment, on the right side. Remove the right cover to access the cartridges. See Figure 11.3.
If installing the cartridges for the first time, skip this step. If replacing the cartridges, disconnect the cartridges
[3] and [4] from the QC connections. First the ones on top and then the ones on the bottom (See Figure 11.1,
see QC connection/disconnection instructions at the beginning of Section 7.3.1).
Connect the new cartridges from the QC connections. First the ones on the bottom and then the ones on top
(see QC connection/disconnection instructions at the beginning of Section 7.3.1).
Open the feed water shut-off valve.
Once the resin cartridges have been replaced, the equipment is put into operation and the conductivity of the
produced water will be shown on the screen.