C
t
C
25
1
. 0
02
t
25
where C
is the water conductivity in µS/cm compensated at 25°C, C
is the water conductivity in µS/cm at a
25
t
temperature t and t the temperature of the water in °C at which the measurement is being carried out.
The Micromatic does not measure the water temperature and, by default, it assumes a water temperature of 25 °C.
If the user knows the water temperature, the user can modify the water compensation temperature value by another
value within a range between 5 and 35 ° C following the steps described in Section 10.4.1. In the main menu, the water
conductivity value will be compensated at 25 ° C.
21/31