PIPETMAN
in the liquid (see immersion depth table).
Release the push-button slowly and
smoothly (to top position) to aspirate the set
volume of liquid.
Wait one second (time depends on model,
see table); then withdraw the pipette-tip
from the liquid.
You may wipe any droplets away from the
outside of the tip using a medical wipe,
however if you do so take care to avoid
touching the tip's orifice.
4) Dispense.
Place the end of the tip against the inside
wall of the recipient vessel (at an angle of
10° to 40°).
Press the push-button slowly and smoothly
to the first stop.
Wait for at least a second; then press the
push-button to the second stop to expel
any residual liquid from the tip.
Keep the push-button pressed fully down
and (while removing the pipette) draw the
tip along the inside surface of the vessel.
Release the push-button, smoothly.
8
P
®
Top
1st Stop
2nd Stop