PIPETMAN
Upper Part (body)
1) Wipe the upper part (body) of the pipette
with a soft-cloth or lint-free tissue
impregnated with the chosen decontaminant.
2) Wipe the upper part of the pipette with a
soft-cloth or lint-free tissue impregnated
with distilled water or sterile water.
Lower Part
The following components only can be im-
mersed in a decontaminant solution: tip-ejector,
tip-holder, connecting nut, piston assembly.
1) Disassemble the pipette as described in
"Maintenance".
2) Immerse the components in the de-
contaminant solution or wipe them according
the instructions given by the manufacturer
or supplier of the decontaminant.
3) Rinse the individual components with
distilled or sterilized water.
4) Leave the parts to dry by evaporation or wipe
them with a clean lint-free tissue or a soft-
cloth.
5) Reassemble the pipette as described in
"Maintenance".

14 - SPECIFICATIONS

Pipetman P is a high quality pipette that offers
excellent accuracy and precision. The figures
given in the "Gilson Maximum Permissible Errors"
table were obtained using Gilson Diamond Tips.
These figures are only guaranteed by using
genuine Gilson Diamond Tips. Refer to Chapter
7, Gilson Diamond Tips.
Each pipette is inspected and validated by
qualified technicians according to the Gilson
Quality System.
The data given in the table conform to the ISO
8655-2 Standard.