Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PIPETMAN
- aspirate the set volume from a beaker of
distilled water,
- maintain the pipette in the vertical position
and wait for 20 seconds,
- if a water droplet appears at the end of the tip
there is a leak,
- if you see no droplet, re-immerse the tip below
the surface of water,
- the water level inside the tip should remain
constant; if the level goes down there is a leak.
P1000 to P10ml
- fit a Gilson Diamond tip,
- set the pipette to the maximum volume given
in the specifications,
- aspirate the set volume from a beaker of
distilled water,
- maintain the pipette in the vertical position
and wait for 20 seconds,
- if a water droplet appears at the end of the tip,
there is a leak.

12 - MAINTENANCE

Routine maintenance will help keep your pipette
in good condition, ensuring a continued high
level of performance. Maintenance is limited to
cleaning or autoclaving the parts specified
under "Cleaning and Decontamination" or to
replacing the push-button, connecting nut, tip-
ejector, tip-holder, seal and O-ring.
Pipetman P2 and P10 should not be dis-
assembled, so you may only replace the push-
button, tip-ejector, dual position tip-ejector and
its adapter; with these pipettes if the tip-holder
is damaged, the piston may also be damaged.