Diode Test

Prior to any diode test, it must be ensured, that the diode to be tested does not
have live voltage across it. Failure to comply with this prescription can lead to
dangerous user injuries or cause instrument damage. Additionally, foreign
voltages falsify the measurement result.
Resistors and semiconductor paths in parallel to the diode cause falsified
measurement results.
1) Position measurement function selection switch to
2) Connect the black test lead to the COM socket and the red test lead to
the
socket.
3) Connect test leads to UUT.
4) Read the measurement result displayed on the screen.

Current Measurement

Ensure that the measurement circuit is not live when connecting the
measurement instrument.
The instruments may only be used in current circuits protected at 16 A up to a
nominal voltage of 600 V.
The nominal cross section of connecting line has to be respected and a safe
connection has to be ensured.
After instrument fuse tripping eliminate the cause for the tripping prior to fuse
replacement.