Assembly and Installation
VI
Remove Calesca from packaging. Allow unit to reach room temperature before
use. Review instructions and assembly. Place sticker with display explanation
in your language on Calesca if needed.
1
Installation on a counter space or other hard surface
2
Optional: Installation on an IV pole
3
7 4
5
8
EN
Connect the power cord provided (choose the correct plug
for your country) with the device in the socket located on the
back of the unit.
Push the wire cord retainer located around the device power
socket down over the plug until secure.
NOTE: For an ideal sound pressure level, we recommend to
place the device 1 meter or more away from the patient.
Make sure that surface is clean and dry. Place the device on
a solid surface. Ensure the device is secure on the surface
to avoid tipping or falling. Place the device on a surface low
enough for staff to see into the device cavity.
Do not block vents on bottom or
rear of device. Do not place on a
soft surface.
Pin not
Pin
engaged
engaged
While securely holding the device in one hand, carefully slide
the U-shaped portion of the adapter into the desired IV slot
located on the device until it clicks into place. Position the
adapter onto the desired IV pole. Make sure the device does
not interfere with any other equipment on the IV pole.
To secure, turn the triangular knob clockwise until the clamp
is tight onto the IV pole.
Do not block vents on
bottom or rear of unit.
To remove the device from the IV pole bracket, pull pin out,
turn ¼ turn, and release. The pin will remain disengaged and
allow for device to slide off the bracket.
IV pole bracket does not interfere with countertop use of
device if left attached.
Remove IV pole adapter from pack-
aging (sold separately). Locate the
two IV positioning slots on either
side of the device. Verify the pin on
the IV pole adapter is engaged (see
picture). If pin is not engaged, turn
the round knob clockwise until the
pin engages.