18
19
NOTE: Calesca was designed to thaw milk from recommended freezer temperatures
(approximately -20˚C/-4˚F). If milk contains any liquid, the final temperature of the milk
at the end of the thawing cycle may fall above the designed range of the device. Milk
should be thawed from a completely frozen state.
Cleaning and Hygiene
VIII
CAUTION:
• Always unplug the appliance from the mains before cleaning it.
• Use drinking water quality for any of the following described cleaning procedures.
• Check with the manufacturer prior to using any cleaning or decontamination method
outside of manufacturer's recommendations.
• Do NOT submerge, autoclave, heat, steam, ETO, or radiation sterilise the warmer.
• Grease on metal hinge may be visible and is considered normal. DO NOT remove
grease as it may shorten the life of the hinge.
• Use of the warmer by more than one user without proper cleaning may pose
a health risk. Inserts are single-patient use items. Use of inserts for more than one
patient may pose a health risk.
• Equipment user has the responsibility for carrying out appropriate decontamination
if hazardous material is spilled on or inside the equipment.
NOTE: These recommendations are not substitutes for official hospital procedures that
may differ among institutions. Refer to hospital policy for cleaning of shared capital
equipment.
Acceptable
cleaning
detergents*
Hygiene
There are two timers located on the LCD display. The lower
left display will show the total cycle time and will count down
to zero. The upper left display shows the total elapsed time
the milk or supplement has been in the device.
If the lid is opened during the thawing cycle, the unit will go
into a pause mode for 60 seconds.The visual indicator on the
front of the unit will flash orange. If the syringe or container is
placed back into the unit, the start button
and the cycle will continue to completion. If the 60 seconds is
exceeded, the unit will stop and reset.
NOTE: Open the lid to activate pause mode while using a
syringe.
When the thawing cycle is complete, the warming action will
automatically shut off. The visual indicator will continue to
flash green and the audible indicator (if not muted) will chime
every 15 seconds. Remove milk immediately while ice crys-
tals are present.
IMPORTANT: If no ice crystals are present, refer to your
hospital breast milk handling and storage policy.
• Cavicide
• PDI
®
• Isopropyl alcohol
• Hydrogen Peroxide 3% liquid or wipes
• Soap and water
*Not responsible for formula changes by
disinfectant wipe manufacturers.
Wash hands before operating the device,
opening inserts, and/or installing inserts.
cleaner and/or wipes
®
Sani-Cloth
cleaner and/or wipes
®
can be pressed,
start
EN
13