Cleaning and Hygiene
IX
Device Care
Disposable Inserts Handling
14
EN
Preparation
Calesca should be cleaned prior to first use and
between patients. This product's inserts are sin-
gle-patient use items and should not be shared.
Use of the device by more than one user without
proper cleaning may pose a health risk. Wash
hands before operating the warmer, opening
inserts, and/or installing inserts.
Between
Remove Medela Disposable Insert and clean with
each patient
an acceptable cleaning agent using wipes or a
use
lint-free cloth. Do not spray or pour liquid directly
into or onto the device.
Allow to air dry before use.
Single
The device should be wiped down with an
patient use
acceptable cleaning agent prior to use and
between patients. If milk is spilled on the device,
remove insert and wipe down unit.
Multiple
If the device is used for more than one patient,
patient use
inserts should stay with the patient with clear and
legible indicators of patient name, date and time.
Inserts are single patient use items. It is recom-
mended that the outer surface of the device is
wiped down with an acceptable cleaning agent
between patients.
In use
Medela Disposable Inserts are a single-patient
and limited use item. Medela recommends
changing the inserts every 12 hours. Inserts
should be changed if human breast milk or
supplement comes in contact with the inserts.
Inserts should also be changed if visible dam-
age has occurred that would offer the potential
for human breast milk or supplement to come in
direct contact with the device.
Insert care
Inspect insert after each warming and/or thawing
session for condensation and/or milk. Inserts are
single patient use items. Remove and discard the
Disposable Insert if patient identified on label no
longer requires feed warming.
If con-
Condensation may appear in the insert if the fro-
densation
zen bottle placed and thawed in the device had
appears in
frost on its outside. Leave the lid to the warmer
insert
open to allow the insert to air dry in between
uses.
If milk
Discard insert immediately. Install a new insert
appears in
prior to next use.
insert