Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and
solutions cause damage to the devices.
• Channels and crevices found on these devices
require particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using hard
water. Instead use purified water unless otherwise
specified.
Limitations on Reprocessing
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH
Detergent
Method
1. Prepare equipment for cleaning by disassembling
removable parts and loosening connections.
2. Rinse equipment in warm water before placing
into washer.
3. Load device into the washer and place small parts
in container or tray inside the washer unit in order
to avoid loosing small components.
NOTE – Load devices carefully into washer in
order to avoid collision.
4. Follow the instructions listed below and set
washer machine to these exact parameters:
Table 2
Phase
Time
Water
(Min.)
Tempera-
ture
Pre-
4:00
Cold water N/A
wash 1
Enzyme
4:00
Hot water
Wash
Detergent and
Concentration
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio 1:128)
10
Table 2
Phase
Time
Water
(Min.)
Tempera-
ture
Wash 1
10:00
60.0ºC
(140º F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2ºC
Rinse**
(180º F)
** Optional phase for disinfection of components
– minimum water temperature as indicated or per
worker manufacturer specifications for the thermal
rinse cycle.
NOTE – Any deviation from this guideline could re-
sult in damage to the equipment as well as improp-
er cleaning results. Rinse with purified water. Do
not perform if parameters cannot be achieved.
5. Remove from washer and dry completely if
needed.
6. Inspect equipment to make sure there is no
visible organic debris or residue from cleaning
agent.
Repeat process if any visible soil is detected.
Inspection
Carefully inspect all parts after each cleaning
to assure all components are in proper working
condition. Pay particular attention to the retractor
arm assemblies.
CAUTION:
Components should be returned for repair if
wear is noted.
Lubrication
It is extremely important that movable parts be
properly lubricated to keep these parts functional. It
is recommended that all components be immersed
in a water-soluble lubricant prior to sterilization.
EN – ENGLISH
Detergent and
Concentration
Neutral pH
detergent
(e.g. Renu-
Klenz™ , Steris,
Ratio 1:256)
N/A
N/A