Cleaning
Prior to cleaning, disassemble the Integra Brain Retractor System into its components as
shown in Figure 1.
Do not disassemble the Retractor Arm assembly or the Handrest assembly. Loosen tension
on each instrument's tension-setting knob, which will allow for ample separation of the
individual ball joint segments for appropriate cleaning.
After each procedure, use a soft brush and mild detergent to remove all residual debris
from each instrument. It is recommended that the instruments and parts be ultrasonically
cleaned.
Lubrication
It is extremely important that movable parts be lubricated to keep these parts functional.
It is recommended that all components be immersed in a water-soluble lubricant prior to
sterilization.
Sterilization
After use, disassemble and clean the Integra Brain Retractor System as described above.
The Integra Brain Retractor System must be cleaned and sterilized prior to each use.
Steam sterilize the Integra Brain Retractor System according to the following validated
sterilization parameters.
Pre-Vacuum Steam Sterilization – Full Cycle:
Sterilization temperature: 132° C
Sterilization exposure time: 8 minutes
It is the user's responsibility to validate any sterilization parameters which deviate from
those listed above.
Follow the sterilizer manufacturer's instructions for operation and loading of steam
autoclave. There must be direct steam exposure to all surfaces of the instruments.
Allow components to cool to room temperature before use.
WARNING:
Do not sterilize the Retractor Arms or the Handrest while the cable is under tension.
Maintenance and Care
To ensure proper function and to extend the life and performance of the equipment, Integra
LifeSciences recommends the following:
Recommended Action
Return the device to the Integra LifeSciences Repairs
department for detailed inspection and servicing.
Request that Integra NeuroSpecialists perform routine
inspections of the device
Recommended Frequency
Once / year
Twice / year
15
EN – ENGLISH