Figure 5.3: The valve is blocked by debris, so the sort channel to the positive collection chamber is blocked. This is not always visible.
The sort channel is clogged
The valve is moving but cells are not moving down the sort channel. This can occur if an air bubble (
5.4 , A) or debris ( Figure 5.4 , B) clogs the sort channel.
Figure 5.4: An air bubble (A) or debris (B) clogs the sort channel to the positive collection chamber.
Do one or all of the following:
Use the priming fixture to wet the channel (refer to the cartidge data sheet).
l
Clean up the sample (e.g. filtration, density gradient, dead cell removal).
l
Use a new cartridge.
l
The input filter is clogged
Sort and event rate decrease immediately during sort run, but transit time of cells is correct. This can occur if:
The sample contains high amount of debris or cell clusters. Clean up the sample (filtration, density gradient,
l
dead cell removal).
Deactivated or non-functional mixer (sedimentation of cells). Check the mixer. Use a new cartridge if necessary.
l
Figure
5
47