Guidance and manufacturer's declaration – electromagnetic immunity – for spirometer (SP10) that are not
LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The spirometer (SP10) is intended for use in the electromagnetic environment specified below. The customer or
the user of spirometer (SP10) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz
to 2.5 GHz
Compliance
Electromagnetic environment - guidance
level
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the spirometer
(SP10), including cables, than the recommended sepa-
ration distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d= 1.17 √P
d= 2.33 √P
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation dis-
tance in metres (m).
80 MHz to 800 MHz
800 MHz to 2.5 GHz