Table 206
Recommended separation distances between portable and mobile RF communications equipment and the PREVENA
PLUS™ 125 Therapy Unit
The PREVENA PLUS™ 125 Therapy Unit is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the PREVENA PLUS™ 125 Therapy Unit can help
prevent electromagnetic interference by maintaining a minimum distance between the portable and mobile RF
communications equipment (transmitters) and the unit as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter in
watts (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
INCLUDED POWER SUPPLIES
Part Number
44000919
44001674
413628
The use of electrical cables and accessories other than those specified in the supplied instructions for
use or referenced documents may result in increased electromagnetic emissions from the PREVENA
PLUS™ 125 Therapy Unit or decreased electromagnetic immunity of the PREVENA PLUS™ 125 Therapy
Unit.
Separation distance according to frequency of transmitter in meters (m)
150 kHz to 80 MHz
d = 1.2 √P
.12
.38
1.2
3.8
12
Description
Power Supply, 240V
Power Supply, 5V, 10W
Cord, V.A.C.® Power, US
80 MHz to 800 MHz
d = 0.35 √P
.04
.11
.35
1.1
3.5
Manufacturer
Inventus Power
Inventus Power
Consolidated Wire
37
800 MHz to 2.5 GHz
d = 0.7 √P
.07
.22
.7
2.2
7
Max Length
3.10 m
1.03 m
2.08 m