CORRECTING A LEAK CONDITION

When the therapy unit detects a significant leak, a visual and audible leak alert will activate (see
Alerts section).

CHECK CANISTER TUBING CONNECTION

1. With the therapy unit on, press down through drape onto the
adhesive seal all the way around to ensure seal.
2. Use fingers to smooth out any creases or wrinkles.
3. Use excess KCI Drape along the outer edge of drape to seal leaks.
1. Ensure canister is securely locked onto the therapy unit. When
canister is installed, a distinct click will be heard indicating it has
been properly installed.
2. Check dressing tubing connector at canister.
3. Check tubing connectors to ensure they are fully engaged and
locked.