CUSTOMER CONTACT INFORMATION
For questions regarding this product, supplies, maintenance or additional information about KCI
products and services, please contact KCI or a KCI authorized representative, or:
In the US call 1-800-275-4524 or visit www.acelity.com
Outside the US visit www.kci-medical.com
KCI USA, Inc. 12930 IH 10 West, San Antonio, TX 78249
PREVENA PLUS™ THERAPY SYSTEM SPECIFICATIONS
Dimensions 60300EP:..................................................................... 3.5" W x 6.4" H x 2.16" D (8.9 x 16.3 x 5.49 cm)
Dimensions 60300P: ....................................................................... 3.5" W x 6.4" H x 1.81" D (8.9 x 16.3 x 4.62 cm)
Weight 60300EP (with empty canister attached): ......................................................................~.64 lbs (~.29 kg)
Weight 60300P (with empty canister attached): .........................................................................~.75 lbs (~.34 kg)
Pressure: ........................................................................................................................................... 125 mmHg (16.7 kPa)
Canister Volume: ........................................................................................................................................................ 150 mL
Electrical:
Battery Run Life: ...................................................................................................................................................~8.5 hours
Battery Charge Time: ................................................................................... ~6 hours from a fully discharged state
External Power Supply Input: ............................................................................. 100-240VAC 0.5A-0.3A 50 - 60 Hz
External Power Supply Output 60300EP: ........................................................................................................5V, 2.0 A
External Power Supply Output 60300P: ..........................................................................................................9V, 2.2 A
Patient and Enclosure Leakage Current: ........................................................................................ <100 Microamps
Environmental Conditions:
Storage/Transport Conditions
Temperature Range: ............................................................................................................ 0°F (-18°C) to 140°F (60°C)
Relative Humidity Range: ...................................................................................................... 0-95% non-condensing
Operating Conditions
Temperature Range: ..............................................................................................................41°F (5°C) to 104°F (40°C)
Relative Humidity Range: .................................................................................................. 15 - 93% non-condensing
Atmospheric Pressure: .........................................1060 hpa (-1253 ft/-381. 9 m) to 700 hpa (9878 ft/3010 m)
Expected Service Life ..............................................................................................................................................7.5 days
IEC Classification
Medical Equipment
Type BF, Applied Part
Class II
IP22 - Protection against solid objects greater than 12.5 mm and against liquid water falling for short
periods of time.
The dressing components of the PREVENA PLUS™ 125 Incision Management System are considered
Applied Parts under IEC 60601-1.
Conforms to AAMI ES60601-1:2005 +A1:2012, IEC 60601-1-6 Ed.3.1:2013, IEC 60601-1-8 Ed.2.1:2012,
IEC 60601-1-11 Ed.2.0:2015, CSA C22.2#60601-1:2014 Ed.3
31