INDICATIONS THAT A LEAK CONDITION HAS BEEN CORRECTED

NOTE: Upon correcting a leak condition, a small delay will occur before the therapy unit senses
the correction and silences the alerts.
The therapy unit will continue the alert until condition is corrected. When leak condition has been
corrected, audible alerts will stop, and visual alerts will turn off.
The PREVENA™ Dressing will be compressed.

BATTERY CHARGING

The PREVENA PLUS™ 125 Therapy Unit is battery-operated to facilitate patient mobility. The battery
charge indicator on the user interface will display three levels of charge.
Full Charge
Full charge (approximately nine hours remain)
Medium charge (approximately two - seven hours remaining)
Low charge. When low charge is indicated approximately two hours of therapy remain. Charge
unit immediately to avoid disruption of therapy.
Dressing compressed - system pressure acceptable.
3
Dressing not compressed - system pressure not acceptable.
X
Return to the Correcting a Leak Condition section to continue
pressure correction steps
Medium Charge
25
Low Charge