10 English
Operation
Guidelines for operation
• This sensor is most reliable at residual chlorine concentrations more
than 0.1 ppm (mg/L). A build-up of sediments/contaminates (e.g.,
biological) on the membrane can interfere or prevent later chlorine
measurements.
• This sensor must not be operated in chlorine free water for more than
one day.
• This sensor must not be exposed to pressure impulses and/or
vibrations from the sample water.
User navigation
Refer to the controller documentation for keypad description and
navigation information.
Configure the sensor
Use the Configure menu to enter identification information and display
options for the sensor and to change options for data handling and
storage.
1. Push the MENU key and select Sensor Setup, Configure.
Option
Description
EDIT NAME
Changes the name that corresponds to the sensor on the
top of the measure screen. The name is limited to
10 characters in any combination of letters, numbers,
spaces or punctuation. The default name is the serial
number of the sensor.