Specifications

Specifications are subject to change without notice.
Specification
Details
Measurement range
0 to 20 ppm
Lower limit of detection
30 ppb (0.030 ppm)
Resolution
0.001 ppm (1 ppb)
pH operating range
4 to 9 pH units
Accuracy
Free chlorine (0 to 10 ppm):
(chlorine concentrations
• ± 3% of the reference test
within ±2 ppm or 20%
pH less than 7.2 (±0.2 pH unit)
(whichever is smaller) of the
calibrated point)
• ± 10% of the reference test
pH less than 8.5 (±0.5 pH unit from the pH at
chlorine calibration)
Total chlorine (0 to 10 ppm):
• ± 10% of the reference test
pH less than 8.5 (±0.5 pH unit from the pH at
chlorine calibration)
• ± 20% of the reference test
greater than 8.5
1
The chlorine sensors are not applicable for applications with consistently low (< 0.1 ppm) or zero chlorine concentrations.
2
The test must be conducted at the analyzer sampling point.
1
2
(DPD) at constant
2
(DPD) at stable
2
(DPD) at stable
2
(DPD) at pH
Specification
Details
Repeatability
30 ppb or 3%, whichever is greater
Response time
Free chlorine: 140 seconds for 90% change (T
at a stable temperature and pH
Total chlorine: 100 seconds for 90% change (T
at a stable temperature and pH
Sampling time
Continuous
Interferences
Free chlorine: Monochloramine, chlorine dioxide,
ozone and chalk deposits
Total chlorine: Chlorine dioxide, ozone and chalk
deposits
Pressure limit
0.5 bar, no pressure impulses and/or vibrations
Flow rate
30 to 50 L/hour (7.9 to 13.2 gal/hour)
40 L/hour (10.5 gal/hour) - optimal
Power requirements
12 VDC, 30 mA maximum (supplied by controller)
Operating temperature
5 to 45 °C (41 to 113 °F)
Storage temperature
Sensor: 0 to 50 °C (32 to 122 °F) dry without
electrolyte
Electrolyte: 15 to 25 °C (59 to 77 °F)
Dimensions
195 mm (7.68 in.)/25 mm (0.98 in.)
(length/diameter)
Cable length/type
1 m
Cable connection
5 pin, M12 connector
Measurement method
Reagentless, electrochemical, amperometric,
three-electrode system
Calibration methods
1-point or 2-point (zero and slope) calibration
Temperature compensation