5
QS1
Main Supply Input
L1 + N
2
2
OR
Switch
L1 + L2
15 Amp
PE
D
POWER ENTRY MODULE
200-240V, 50/60 Hz
(Inlet C20, embedded
+/- 10%
Circuit Breaker TA35)
Mains cables:
C19 --- Schuko (1,5mm2)
C19 --- NEMA 6-15 (AWG14)
A2
Secotom-6...1100W
FREQUENCY INVERTER
24VDC OUTPUT P24
(max 1A)
Temp. masur.
input
Power
A2
output
3 x PT100
C
2
M1
3
3
D
MAIN MOTOR
Secotom-6...750W
A5
Y-TABLE MOVEMENT UNIT
4
24VDC
16,5W
B
Stepper Motor Unit
Steppermotor
Address 1
A
5
4
A1
2
POWER SUPPLY
24VDC / 6.3A
Z1
2
2
SB1.2
EMC FILTER
with low leakage current
Safe STOP
STO
Input
YS1
SOLENOID INTERLOCK
Safe STOP
DI3
Input
CAN bus
RJ45
DQ
Door Guard
YS1.2
Lock
(shown
KS1
unlocked)
YS1.2
SS1
Magnetic
switch
3
5
Voltages related to the safety features:
24V-1: Unprotected voltage from 24VDC power supply
24V-2: Protected by Emergency Stop only
24V-3: Protected by Emergency Stop and Door switch
4
3
2
X1
X2
24V-1 input
Output, 24VDC
(unswitched)
(24V-2 chain)
SB1.1
Emergency
Stop
X2
24V-2 input.
(Emergency Stop switched)
X2
YS1.4
Door Guard
Switch
X2
24V-3 input
(Emergency Stop + Door closed)
(shown with
K3
door open)
X4
2
Output, 24VDC / 500mA
(24V-1 chain)
2
Control
Water pump
enable
X14
S1.2
BUS JUNCTION UNIT
5
CAN BUS Interface
X8
(24V-2)
3
2
SB1.3
strap for powering
of M2 motor from
KS1.1
24V-2 voltage
2
S1.2
Water pump
enable
X4
X24
X26
X18
Display Driver
Power Input for
Pump Output
3 x Hall Sensors
Epson
Motor Driver
3 x +/- 0-24V, 5A
Feedback input
S1D13706
(from 24V-2 chain)
(24V-2 chain)
3 x dig. inputs 5V
Backlight Output
12V PWM
3 x Analog Input 5V,
5VDC Supply for joystick
A4
Matrix Keyboard input,
Variant Detect input
Printed Circuit Board
(2 out / 4 in)
(same for both Secotom - 20 and 30)
CPU, V850ES/SJ3, 1M FLASH, 60k RAM
SRAM 512k x 16, NVRAM 32k x 8
RESET GENERATOR
3 x dig. input 5V,
Supply for encoder
REAL TIME CLOCK + BACKUP
PCB TEMPERATURE MONITOR
HEART BEAT INDICATOR
PCB VERSION IDENTIFICATION
VOLTAGE MONITORS
X21
X22
Output, 24VDC / 500mA
(24V-1 chain)
CONN. FOR INITIAL
CONN. FOR DEBUG
PROGRAMMING
X20
BUZZER
CONN. FOR STRUERS
MEMORY MODULE
rev. B (E GA ) - corrected Drawing No.
Friday, June 21, 2019
2
1
M2
DC MOTOR
24V / 12W
PERISTALTIC
D
PUMP
A3
parallel
TFT Color Display,
320x240 dots with
white LED backlight
2
X9
SB2
6
MATRIX
KEYBOARD
C
(6 keys)
BR1
6
TURN-PUSH
BUTTON
HL1
2
Working
light
B
4
X51
USB
Interface
CONNECTOR FOR SERVICE
A
S TRUE RS A/S
Pederstrupvej
84
DK -2750 Ballerup
Denmark
Secotom - 6, Block Diagram
Size
CA GE Code
DW G NO
Rev
16853051
B
A 2
Sheettt
Scale
EGA
1
of
1
1